4-12
IM DL850E-01EN
AND Trigger (Enhanced)
The DL850E/DL850EV triggers on the AND of multiple trigger source conditions. The DL850E/DL850EV triggers
when all the specified conditions are met at a single point.
Trigger Source
You can use CH1 to CH16 and Bit 1 to Bit 8 as the trigger sources. Bit 1 to Bit 8 appear as options when a logic
input module is installed in a slot.
Achievement Condition (Condition)
Set the achievement condition for each trigger source.
H
The signal level must be high.
L
The signal level must be low.
IN
The signal must be within the specified level range.
OUT
The signal must be outside of the specified level range.
−
The signal is not used as a trigger condition.
* IN and OUT are selectable only when the trigger source is an analog signal (CH1 to CH16).
Trigger Level (Level)
Set this item for each trigger source.
When the Achievement Condition Is H or L
Set the level for determining whether the trigger sources are high or low.
When the Achievement Condition Is IN or OUT
An edge is detected when the trigger source enters (IN) or leaves (OUT) the specified level range. You can
specify the level range settings for each analog signal trigger source.
Trigger Hysteresis (Hys)
Set this item for each trigger source.
This item is the same as the hysteresis
of the simple trigger.
Trigger Hold-Off (Hold Off), Trigger Position (Position), Trigger Delay (Delay)
These items are the same as trigger hold-off
, trigger position
4 Triggering