IM 701450-61E

3

Turning ON the Power Analysis Function/
Selecting the Attenuation or Current-to-Voltage
Conversion Ratio of Probes/Enabling Waveform
Computation Setup

Procedure

•  To exit the menu during operation, press 

ESC

 located 

above the soft keys.

•  In the procedural explanation below, the term 

jog shuttle & SELECT

 refers to the operation of 

selecting/setting items and entering values using the 

jog 

shuttle

SELECT

 and 

RESET

 keys.  For details on the 

operation using the jog shuttle, SELECT, and RESET, see 
sections 4.1 or 4.2 in the 

DL7440/DL7480 User’s Manual

.

•  For a description of the operation using a USB keyboard 

or a USB mouse, see section 4.3 in the 

DL7440/DL7480 

User’s Manual

.

X-Y

PHASE

MENU

MENU

SETUP

FILE

MISC

RESET

SELECT

MATH

SHIFT

GO/NOGO

CURSOR

MEASURE

HISTORY

PRINT

ACQ

START/STOP

IMAGE  SAVE

DISPLAY

1.

Press 

SETUP

.  The SETUP menu appears.

2.

Press the 

Power Analyze Setup

 soft key. The Power Analyze Setup dialog box opens.

Turning ON the Power Analysis Function

3.

Use 

jog shuttle & SELECT

 to select ON or OFF for each power analysis

channel pair (PWR1 Analyze, PWR2 Analyze, and PWR3 Analyze (PWR3

Analyze is available only on the DL7480)).

When OFF is selected, power analysis will not be performed on the channel pair.

When ON is selected, power analysis will be performed on the channel pair.

Turn ON/OFF the power
analysis function

Power analysis is performed
on a pair of channels that
are turned ON.

Selecting the Probe Attenuation or Current-to-Voltage Conversion Ratio

4.

Use 

jog shuttle & SELECT

 to select the attenuation of the voltage probes of the

voltage input channels (CH1, CH3, and CH5 (CH5 is available only on the DL7480)).

5.

Use 

jog shuttle & SELECT

 to select the current-to-voltage conversion ratio of

the current probes of the current input channels (CH2, CH4, and CH6 (CH6 is

available only on the DL7480)).

Select the probe attenuation/current-to-
voltage conversion ratio

Select the attenuation of the voltage probe of 
the voltage input channel and the current-to-
voltage conversion ratio of the current probe 
of the current input channel.

Содержание DL7440

Страница 1: ...User s Manual Yokogawa Electric Corporation DL7440 DL7480 Digital Oscilloscope Power Supply Analysis Function G4 Option User s Manual IM 701450 61E 2nd Edition...

Страница 2: ...f the DL7400 Digital Oscilloscope that supports the Power Supply Analysis Function G4 Option is 1 20 or later For instructions on checking the firmware version see section 16 4 in the DL7440 DL7480 Us...

Страница 3: ...steps Explanation This subsection describes the setup parameters and the limitations on the procedures Notations Used in the Procedures Panel Keys and Soft keys Bold characters used in the procedural...

Страница 4: ...l Processing on the Measured Values of Power Analysis Parameters 17 7 Performing Waveform Computation on Power Analysis Parameters 18 8 Displaying the Trend of the Measured Values of Waveform Paramete...

Страница 5: ...tude UP P maximum value U pk minimum value U pk DC component Udc rms value Urms AC component Uac rectified mean value calibrated to the rms value Umn and rectified mean value Urmn Current Amplitude IP...

Страница 6: ...of 600 W or less Class D Class C Portable tools Lighting equipment Balanced three phase equipment Household appliances excluding equipment identified as Class D Tools excluding portable tools portabl...

Страница 7: ...g of voltage probes see the manual that came with the product For instructions on the phase correction of probes see section 3 5 in the DL7440 DL7480 User s Manual Degaussing Current Probes and Perfor...

Страница 8: ...box opens Turning ON the Power Analysis Function 3 Use jog shuttle SELECT to select ON or OFF for each power analysis channel pair PWR1 Analyze PWR2 Analyze and PWR3 Analyze PWR3 Analyze is available...

Страница 9: ...rs using the power analysis function G4 option you must turn ON the power analysis function select the voltage probe attenuation and select the current to voltage conversion ratio of current probes In...

Страница 10: ...the DL7400 and select a current to voltage conversion ratio of 10A 1V the DL7400 displays the current value measured on the current probe as 10 A when the output voltage from the current probe is 1 V...

Страница 11: ...L7400 as shown below For a description on the pair of channels for applying the voltage and current signals when measuring power analysis parameters see page 6 in this manual 3 Set the attenuation for...

Страница 12: ...the Auto Deskew Execute soft key Deskew is automatically executed Executing Manual Deskew You can also manually execute deskew You can use manual skew to correct the difference further after executing...

Страница 13: ...el Target CH to match the signal of the channel set to be the trigger source of edge trigger along the time axis It is a function used to correct the difference in the transfer time Execute auto deske...

Страница 14: ...y selecting To Measure in the Power Analyze Setup dialog box described in section 3 of this manual and pressing SELECT 2 Press the Mode soft key The Mode menu appears 3 Press the ON soft key Selecting...

Страница 15: ...the procedure is the same as steps 9 to 32 in section 10 6 pages 10 47 to 10 49 in the DL7440 DL7480 User s Manual IM701450 01E Performing Automated Measurement of Power Analysis Parameters on Dual A...

Страница 16: ...ac Umn Urmn S P Q Z Wp Wp and Wp Standard measurement parameters For details on how to determine each parameter section 10 6 in the DL7440 DL7480 User s Manual OShot OShot Freq Period Rise Fall Width...

Страница 17: ...Iac I pk IP P DC component Idc True rms value Irms AC component Iac Rectified mean value calibrated to the rms value Imn Rectified mean value Irmn Amplitude IP P Maximum value I pk Minimum value I pk...

Страница 18: ...ure in the Power Analyze Setup dialog box described in section 3 of this manual and pressing SELECT The rest of the procedure is the same as steps 2 to 15 in section 10 7 pages 10 54 to 10 58 in the D...

Страница 19: ...lay Math1 Display or Math2 Display and the procedure for setting computed waveform labels Math1 Label or Math2 Label see section 9 1 in the DL7440 DL7480 User s Manual To make correct measurements and...

Страница 20: ...Set the computation start point When the operator is Power and the computed waveform is CH5 or CH6 the maximum record length that can be computed is from the computation start point to 2 MW 6 Press E...

Страница 21: ...the Computed Waveform of Power Spectrum When PS Was Selected in Step 4 on Page 18 5 Use jog shuttle SELECT to set the computation source waveform scaling unit computation start point and time window o...

Страница 22: ...cable class A or B The harmonic measurement data and the standard limits for each order are listed When the Applicable Class of the Equipment under Test Is C Select Harmonics Select the waveform on wh...

Страница 23: ...graph Waveform to be computed Limit displayed in white on the screen Computed value of harmonics List Display Example The harmonic measurement data and the standard limits for each order up to the 40t...

Страница 24: ...Be Computed The waveforms on which computation can be performed source are as follows Computation Name Operator Source Source1 Source2 Math1 Power CH1 CH2 CH5 CH6 Z CH1 CH2 CH5 CH6 I2 t CH2 CH6 PS CH...

Страница 25: ...Normal For a description of the trigger mode see section 6 1 in the DL7440 DL7480 User s Manual Time window Rect Rectangular Number of waveforms and number of waveform data points To perform computat...

Страница 26: ...s D the harmonic current per watt power ratio limit is also evaluated Displaying the computed results Bar graph display The harmonic measurement data and the standard limits for each order up to the 4...

Страница 27: ...the measurement of waveform parameters per cycle stop waveform acquisition Turn ON the computed waveform display For the setup procedure see section 9 1 in the DL7440 DL7480 User s Manual Note The se...

Страница 28: ...rement is displayed in yellow on a black background If the characters are in yellow but the background is not black it is only selected as a measurement parameter The parameter does not have measured...

Страница 29: ...ine each parameter section 10 6 in the DL7440 DL7480 User s Manual OShot OShot Freq Period Rise Fall Width Width Duty Burst1 Burst2 Pulse High Low AvgFreq AvgPeriod Int1TY and Int2TY CH7 CH8 Math1 and...

Страница 30: ...0 User s Manual IM701450 01E Selecting the Search Target Waveform and Search Measurement Parameter 8 Press the Item Setup soft key The Item Setup menu and Item Setup dialog box appear 9 Press one of t...

Страница 31: ...ecified on the searched waveform CH7 CH8 Math1 Math2 and CH1 to CH6 when power analysis is not specified on the searched waveform CH5 to CH8 can be used only on the DL7480 For CH5 to CH8 measurement p...

Страница 32: ...ls on how to determine each parameter section 10 6 in the DL7440 DL7480 User s Manual OShot OShot Freq Period Rise Fall Width Width Duty Burst1 Burst2 Pulse Delay delay between waveforms High Low AvgF...

Страница 33: ...on 4 in this manual 1 Press GO NOGO The GO NO GO menu appears The rest of the procedure is the same as steps 2 to 13 in section 10 10 pages 10 72 to 10 74 in the DL7440 DL7480 User s Manual IM701450 0...

Страница 34: ...ECT and RESET keys For details on the operation using the jog shuttle SELECT and RESET see sections 4 1 or 4 2 in the DL7440 DL7480 User s Manual For a description of the operation using a USB keyboar...

Страница 35: ...n Math1 is Class A and Math2 is not harmonic computation 3129 bytes minimum The data size vary between 3192 and 9157 bytes depending on the settings Extension The extension is CSV Note This function c...

Страница 36: ...the power factor for Class C or queries the current setting 42 MATH x HARMonics LIST Queries the computed values of harmonics and limits defined by the standard for 42 each order MATH x HARMonics MAX...

Страница 37: ...ower analysis 47 PANalyze MATH x MODE Enables Disables the computed waveform MATH x of power analysis or queries 47 the current setting PANalyze PWR x Queries all settings related to the power analysi...

Страница 38: ...Cute ABORt FILE SAVE HARMonics ABORt Function Aborts the saving of the computed results of harmonics Syntax FILE SAVE HARMonics ABORt Example FILE SAVE HARMONICS ABORT FILE SAVE HARMonics EXECute Func...

Страница 39: ...d the upper and lower limits or queries the current setting Syntax GONogo PARameter ITEM x TYPE Parameter Voltage DONTcare Current DONTcare Current DONTcare Current DONTcare NRf DONTcare NRf DONTcare...

Страница 40: ...ower limits or queries the current setting Syntax HISTory PARameter ITEM x TYPE Parameter Voltage DONTcare Voltage DONTcare Current DONTcare Current DONTcare NRf DONTcare NRf DONTcare HISTory PARamete...

Страница 41: ...s Space A B C D LAMBda Space NRf LIST Space NRf MAXCurrent Space OPOWer Space OFF ON NRf POWer Space NRf SPOint Space NRf VOLTage Space OPERation Space POWer NRf NRf NRf NRf Z I2T PPS HARMonics TRENd...

Страница 42: ...SS C OPOWER 1 POWER 1 00000E 00 LAMBDA 80 0 DMODE LINEAR MATH x HARMonics CLASs Function Sets the applicable class of the equipment under test or queries the current setting Syntax MATH x HARMonics CL...

Страница 43: ...ENT 50 000 MATH x HARMonics OPOWer Function Sets whether active power of 25 W is exceeded or queries the current setting Syntax MATH x HARMonics OPOWer Boolean MATH x HARMonics OPOWer x 1 to 2 Example...

Страница 44: ...NRf MATH x PFFT SPOint x 1 to 2 NRf 5 to 5 The resolution is 10 div display record length Example MATH1 HARMONICS VOLTAGE 220 MATH1 HARMONICS VOLTAGE MATH1 HARMONICS VOLTAGE 220 000000E 00 MATH x PFFT...

Страница 45: ...and statistical processing on power analysis parameters MEASure CHANnel x AH COUNt SDEViation MAXimum MEAN MINimum STATe Space OFF ON NRf VALue AHN AHP I2T IAC IDC IMN INPeak IPPeak IPTopeak IRMN IRM...

Страница 46: ...mple MEASURE CHANNEL1 UDC STATE ON MEASURE CHANNEL1 UDC STATE MEASURE CHANNEL1 UDC STATE 1 Description The power analysis parameters below can be substituted using standard waveform parameters UPTopea...

Страница 47: ...commands in this group deal with settings of the power analysis function PANalyze MATH x MODE Space OFF ON NRf PWR x I PROBe Space MODE Space OFF ON NRf C10 C100 MATH U PROBe Space NRf MEASure JUMP Sp...

Страница 48: ...Function Enables Disables the power analysis target PWR x or queries the current setting Syntax PANalyze PWR x MODE Boolean PANalyze PWR x MODE x 1 to 3 1 or 2 on the DL7440 Example PANALYZE PWR1 MOD...

Страница 49: ...owing cases When the trigger type is not Simple When the trigger type is Simple and the source is Ext or Line When Trigger Source Deskew Target CH When Deskew Target CH is Pod A or Pod B 45 Auto deske...

Страница 50: ...lues of power analysis parameters Waveform computation on As with the standard waveform computation performs waveform computation such as power analysis parameters Instantaneous power impedance Joule...

Страница 51: ...ation is 1000 1 Channel with current probe connected 20 0 mA div TIME DIV Time axis 20 ns div MODE Trigger mode When correction is executed and is successful Normal DL7400 in start condition When corr...

Страница 52: ...13333 5333 2667 1333 533 267 1M 50Hz 640000 320000 160000 64000 32000 16000 6400 3200 1600 640 60Hz 533333 266667 133333 53333 26667 13333 5333 2667 1333 533 2M 50Hz 1600000 640000 320000 160000 6400...

Страница 53: ...current to voltage conversion ratio 9 current to voltage conversion ratio 9 D data size 34 degauss 6 deskew 4 10 deskew correction signal source 4 deskew correction signal source connection of 12 des...

Страница 54: ...GO determination 32 power analysis parameters history search on 29 power analysis parameters statistical processing 17 power analysis parameters waveform computation on 18 Power Analyze Setup 7 power...

Страница 55: ...HARMonics DMODe 41 MATH x HARMonics LAMBda 42 MATH x HARMonics LIST 42 MATH x HARMonics MAXCurrent 42 MATH x HARMonics OPOWer 42 MATH x HARMonics POWer 42 MATH x HARMonics SPOint 42 MATH x HARMonics...

Отзывы: