14 Specifications
Item
Specification
Correction of the difference
Corrects (deskew) the difference in the transfer time of voltage and current signals automatically or
in the transfer time
manually
(Deskew)
The correction range is
±
100 ns (0.01 ns resolution).
Automated measurement of As with the standard measurement parameters (waveform parameters), performs automated
power analysis parameters
measurement of power analysis parameters (see page 4).
Automated measurement on dual areas is also possible.
Statistical processing on the As with the standard measurement parameters, performs statistical processing on the measured
measured values
values of power analysis parameters.
Waveform computation on
As with the standard waveform computation, performs waveform computation such as
power analysis parameters
Instantaneous power, impedance, Joule integral, power spectrum, and harmonics
For waveform computation of harmonics, simple comparison against the limits of IEC 61000-3-2
Edition 2.1, and EN61000-3-2 Amendment 14 is possible.
Trend display
Displays the trend of the change in the measured values of waveform parameters per cycle over
time
History search
As with the standard measurement parameters, performs history search using power analysis
parameters.
GO/NO-GO determination
As with the standard measurement parameters, performs GO/NO-GO determination using power
analysis parameters.
Saving of the computed
Saves the computed result of harmonics to a file in CSV format
results of harmonics.