3-4
IM 701830-01E
3.3
Connecting the Power Cord
Before Connecting the Power
Make sure that you observe the following points before connecting the power. Failure to do so
may cause electric shock or damage to the instrument.
WARNING
• Connect the power cord after confirming that the voltage of the power supply
complies to the rated electric power voltage for the instrument.
• Connect the power cord after confirming that the instrument power switch is OFF.
• Always use protective earth to prevent electric shock. Connect the instrument
power cord to the 3-pin power outlet with earth terminal.
• Do not use non-earth extension cords or other measures that defect the protective
earth.
• Never use an extension cord that does not have a protective earth, otherwise the
protection feature will be invalidated.
Connecting the Power Cord
1.
Make sure that POWER switch is OFF.
2.
Connect the plug of the accessory power cable to the power connector on the left side of the
instrument.
3.
Plug the other end of the power cord into an AC outlet that meets the following conditions.
The AC outlet must be of 3-pin type with a protective earth terminal.
Item
Rated supply voltage
100 to 120 VAC/200 to 240 VAC (Switchable)
Permitted supply voltage range
90 to 132 VAC/180 to 264 VAC
Rated supply voltage frequency
50/60 Hz
Permitted supply voltage frequency range
48 to 63 Hz
Maximum power consumption
250 VA
(when the built-in printer is used)