15-7
15
IM 701830-01E
Specifications
15.16 General
Item
Specifications
Standard operating conditions
Ambient temperature : 23
±
5
°
C
Ambient humidity
: 55
±
10% RH
Power voltage and frequency: Less than 1% of the rated voltage/frequency fluctuation:
After 30-min or more warm up
Storage conditions
Temperature
: -20 to 60
°
C
Humidity
: 20 to 85% RH (no condensation allowed)
Operating conditions
Temperature
: 5 to 40
°
C
Humidity
: 20 to 80% RH (without a printer)
35 to 80% RH (with a printer)
Storage altitude
3000 m or below
Operating altitude
2000 m or below
Rated supply voltage
100 to 120 VAC/200 to 240 VAC (switchable)
Permissible supply voltage range
90 to 132 VAC/180 to 264 VAC
Rated supply voltage frequency
50/60 Hz
Permissible supply voltage frequency48 to 63 Hz
Maximum power consumption
250 VA
(using printer)
Withstand voltage
1.5 k VAC for 1 minute
(between power supply and earth)
Insulation resistance
10 M
Ω
or more at 500 VDC
(between power supply and earth)
External dimensions
Approx. 355 (W)
×
260 (H)
×
305 (D) mm
(details on page 15-8)
(with printer cover closed, handles and projections excluded)
Weight
Main instrument only (701830) : Approx. 9.2 kg
High-Speed Isolation Module
: Approx. 200 g
Unternal Hard Disk (optional)
: Approx. 500 g
Cooling method
Forced air cooling, air discharged from left side and bottom.
Installation position
Horizontal (when a stand is used) or vertical, stacking is not allowed.
Battery back-up
Set-up data and internal clock are backed up by a built-in lithium battery.
Battery life : Approx. 5 years (at ambient temperature of 23
°
C)
Fuse*
location
max. rated voltage
max. rated current
type
standard
Power unit
250 V
5 A
time lag
SEMKO
Accessories
• 1 power cord
• Printer connecting cable (B9946YY)
• Cover plates (16 plates)
• Front cover
• 1 roll of printer paper (Only on models with “/B5” suffix)
• 2 rubber pads, for rear legs
• Soft case
• User’s manual (this manual)
• Operation Guide
• Communication Interface Manual
* Because the fuses used by this instrument is inside the case, they cannot be exchanged by the user. If you believe the fuse inside the case
is blown, please contact your nearest YOKOGAWA dealer listed on the back cover of this manual.