17
IM XL120P
Saving Data
Measured data, calculated data, setting data, and so forth can be saved to the
XL100 internal memory or an external storage medium (CF card or SD card)
that is inserted in the XL100.
SD
CF
Save the data
Supported only on the
XL121, XL122, and XL124
TIP
The XL100 is equipped with a USB port for USB memories. However, data cannot
be saved directly to a USB memory (see section 3.9 in the User’s Manual).
The types of data that can be saved are as follows:
Type
Description
Logging data
The instantaneous values of the measured/calculated data
can be saved at a specified sampling interval. The data save
operation is started or stopped with the START/STOP key.
The save operation can also be started or stopped when a
specific event (see “Trigger” on the next page) occurs. The
logging data contains alarm information.
Manual sampled data
The measured/calculated data (instantaneous values) of all
channels can be saved by pressing the SAVE key in Free
Running Mode.
Alarm data
The same information as the alarm summary display can be
saved by pressing the SAVE key during alarm summary
display.
Screen image data
The image data of the screen being displayed can be saved
by pressing the SAVE key in Free Running Mode, etc.
Setting data
The setting data of the XL100 can be saved in File Operation
Mode.
Log data
The same information as the log display can be saved by
pressing the SAVE key during log data display.
Backup file
If the data save operation is not carried out normally to the
internal memory or external storage medium (CF card or SD
card), the data is saved to the backup memory of the XL100.
The saved data can be copied to an external storage
medium.