Procedure
• Viewing the Calibration Notification Screen
The
calibration notification screen only appears over operation screens. It will continue
to appear until you perform calibration (Finish Calibration).
You can close the calibration notification screen by pressing
ESC
or
DISP/ENTER
.
Operation Example
When Notification is set to 1 (DayBfr) and Renotification is set to 8h
Notification screen
display start
Calibration due date
Notification screen
appears
Renotification
Renotification
Renotification
Renotification
00:00
08:00
16:00
00:00
08:00
Notification (DayBfr) setting
Notification screen
appears
Notification screen
appears
Notification screen
appears
Note
• The calibration notification screen will close if you change screens while it is displayed.
• If the calibration due date passes before you complete calibration, that information will be
recorded in the login log (the operation log on DXs with the /AS1 option).
3.13 Managing the Input Calibration Interval (/CC1 option; release numbers 4 and later)