5-26
IM CW500-01EN
Harmonics
Select whether to record the harmonic data of voltage, current, and
power.
Settings
Record*/Not record
* Default value
Event
Select whether to record detailed data when a power quality event
occurs. If the current range is set to AUTO, “Record” cannot be selected.
To select “Record,” change the current range to a fixed range first.
* If AUTO range is selected, measurements according to international
power quality standard IEC61000-4-30 Class S cannot be performed.
Settings
Record*/Not record
* Default value
<Procedure>
Use the up and down arrow keys to move to Harmonics or Event.
→
Press ENTER to show a pull-down menu.
→
Use the up and down arrow
keys to select Record or Do not record.
→
Press ENTER to confirm or
ESC to cancel.
5.4 Recording Setup