Precautions for Safe Use of the Instrument 

When handling the instrument, ALWAYS observe all of the cautionary notes on 
safety given below. Yokogawa M&C Corporation is not at all liable for damage 
resulting from misuse of this product by the user that is contrary to these cautionary 
notes. 

 

The instrument and this manual use the following safety symbols: 

Danger! Handle with Care. 
This symbol indicates that the operator must refer to an explanation in
the instruction manual in order to avoid the risk of personal injury or
death and/or damage to the instrument. 

 

Double Insulation 
This symbol indicates double insulation. 

 

AC Voltage/Current 
This symbol indicates AC voltage or current 

 

Ground 
This symbol indicates ground (earth) 

 

WARNING 

Indicates that there is a possibility of serious personal injury or loss of life if the 
operating procedure is not followed correctly and describes the precautions for 
avoiding such injury or loss of life. 

CAUTION 

Indicates that there is a possibility of serious personal injury of damage to the 
instrument if the operating procedure is not followed correctly and describes the 
precautions for avoiding such injury or damage. 

NOTE 

Draws attention to information essential for understanding the operation and 
features. 

 

 
 
 
 

WARNING 

  Never make measurement on a circuit above 300V AC. 

  Do not use the instrument in an atmosphere where any flammable or explosive 

gas present. 

  Avoid using the instrument if it has been exposed to rain or moisture or if your 

hands are wet. 

  Do not exceed the maximum allowable input of any measurement range. 

  As to current measurement of more than 1000A, do not make measurement 

continuously. 

  The main unit is heated, so there is danger which damages safety. 

  Do not use the instrument if there is any damage to the casing or when the 

casing is removed. 

  Do not install substitute parts or make any modification to the instrument.   

 

WARNING 

To avoid damage to the instrument or electric shock! 
The restrictions on the maximum voltage level for which the 99025 testers can be 
used, depend on the over-voltage categories specified by the safety standards. 
These category specifications are formulated to protect operators against transient 
impulse voltage in power lines. 

 

Maximum Allowable Input 

Function 

OVERVOLTAGE CATEGORY

 

III 

AC 3000A rms 

Measuring circuit voltage : AC 300V rms 

Over-voltage category I (CAT.I): 

Signal level, special equipment or parts of equipment, telecommunication, electronic etc., with 
smaller transient over-voltages than CAT.II. 

Over-voltage category II (CAT.II) 

Local level, appliance, portable equipment etc., with smaller transient over-voltages than CAT.III. 

Over-voltage category III (CAT.III): 
  Distribution level, fixed installation, with smaller transient over-voltages than CAT.IV. 

 

CATION 

Do not expose the instrument to the direct sun, extreme temperatures or dew fall. 

 

NOTE 

  If equipment generating strong electromagnetic interference is located nearly, 

the testers may malfunction. 

1. Features 

Model 99025 are designed to increase the measuring capability of your clamp-on 
testers. With the use of the Clamp Adapter, you can not only extend current 
ranges, but clamp on a conductor of a larger diameter. 
 

2.  How to Use 

(1)  Clamp your clamp-on tester on the pick-up coil of the Clamp Adapter.    Then 

clamp the Clamp Adapter on the conductor as shown bellow. 

(2)  When you use Clamp Adapter for power or power factor measurement; 

Clamp your clamp-on tester on the pick-up coil of the Clamp Adapter, 
assuming that the arrow mark on the pick-up coil points the load side. 

 

Then clamp the Clamp Adapter on the conductor so that the arrow mark on 
the Clamp Adapter jaw points the load side of the conductor. 

Transformer Jaws

Conductor

Arrow Mark

Arrow Mark

Clamp Adapter

Pick-up Coil

Clamp-on Tester

 

 

3. SPECIFICATIONS 

Measuring Range 

0 to 3000A AC 

Input / Output 

10 : 1 

Accuracy 

±

2% of input 

±

0.5A 

Withstand Voltage 

3700V AC for one minute 

Dimensions 

Clamp Adapter except Pick-up coil 
150(W) X 317(H) X 33(D)mm 
Pick-up coil 

 

45(W) X 40(H) X 10(D)mm 

Weight Approx. 

750g 

Conductor Size 

100mm max. Diameter 
100 X 150mm Bus-Bar 

Frequency 

50Hz / 60Hz 

Duty Cycle 

Continuous for 

0 to 1000A 

10 minutes for 

1000 to 1500A 

30 seconds for 

1500 to 3000A 

Safety Standards 

EN 61010-1 

 EN 

61010-2-032 

 

AC 300V CAT. III / AC 600V CAT. II, Pollution degree2 

Accessories 

Carrying Case (Model 93035) 

 
After-sales Service 
Should any failure occur while you are using the tester, follow the instructions 
given below.  If the adapter still fails to operate correctly and needs repair, 
contact the vendor from whom you purchased the instrument or the nearest 
Yokogawa M&C sales office.   

 

YOKOGAWA M&C CORPORATION 
International Sales Dept. 

2-9-32 Nakacho, Musashino-shi, Tokyo, 180-8750 Japan 
Phone: 81-422-52-5716  Facsimile: 81-422-55-8654 

YOKOGAWA CORPORATION OF AMERICA (U.S.A.) 

Phone: 1-770-253-7000  Facsimile: 1-770-251-2088 

YOKOGAWA EUROPE B. V. (THE NETHERLANDS) 

Phone: 31-334-64-1611  Facsimile: 31-334-64-1610 

YOKOGAWA AMERICA DO SUL S. A. (BRAZIL) 

Phone: 55-11-5681-2400    Facsimile: 55-11-5681-1274 

YOKOGAWA ENGINEERING INSTRUMENTS KOREA CORPORATION (KOREA) 

Phone: 82-2-551-0660 to -0664    Facsimile: 82-2-551-0665 

YOKOGAWA AUSTRALIA PTY. LTD. (AUSTRALIA) 

Phone: 61-2-9805-0699  Facsimile: 61-2-9888-1844 

YOKOGAWA BLUE STAR LTD. (INDIA) 

Phone: 91-80-227-1513  Facsimile: 91-80-227-4270 

LTD. YOKOGAWA ELECTRIC (RUSSIAN FEDERATION) 

Phone: 7-095-737-7868  Facsimile: 7-095-737-7869 

計 測 機 器 営 業 部

 

180-8750 

東京都武蔵野市中町

2-9-32 

電話:

0422-52-5984 

ファクシミリ:

0422-55-8953 

西

 

564-0063 

大阪府吹田市江坂町

1-23-101 

大同生命江坂ビル

10

 

電話:

06-6368-7041 

ファクシミリ:

06-6368-7045 

 

450-0003 

名古屋市中村区名駅南

1-27-2 

日本生命笹島ビル

12

 

電話:

052-581-7490 

ファクシミリ:

052-581-7664 

 

730-0037 

広島市中区中町

8-12 

広島グリーンビル8階

 

電話:

082-240-7676 

ファクシミリ:

082-541-4567 

 

812-0037 

福岡市博多区御供所町

3-21 

大博通りビジネスセンター

7

 

電話:

092-262-5740 

ファクシミリ:

092-262-5741 

 

180-5679 

東京都武蔵野市中町

2-9-32 

電話:

0422-52-5679

ファクシミリ:

0422-51-8455 

 

KIM1JE-2003.2

 

User’s 

Manual

 

99025 
Clamp Adapter 

IM 99025

IM 99025

1st Edition: Apr.2003(MC)

Отзывы: