3.4
Turning the Power Switch ON/OFF
Items to Be Checked before Turning ON the Power
• Check that the instrument is installed properly (see section 3.2, “Installation”).
• Check that the power cord is connected properly (see section 3.3, “Connecting the
Power Supply”).
Location of the Power Switch and ON/OFF Operation
The power switch is located in the lower left corner of the front panel. The power switch
is a push button. Press the button once to turn it “ON” and press it again to turn it “OFF.”
OFF
ON
SOURCE
MENU
MODE
KEY LOC
K
VALU
E
MEASUR
E
COMPAR
E
AUTO
RANGE
REGAL
L
MATH
NULL
LOCAL
POWE
R
MISC
TIME
SHIFT
SOURCE
MODE
AUTO
RANG
E
LIMIT
VS IS
RANGE
OUTPU
T CONTROL
SWEEP
START
ZERO
TRIG
OUTPUT
V I
Power ON Operation
When the power switch is turned ON, the self-test starts automatically. When the self-
test completes successfully, the GS610 is set to the factory default settings or the
settings corresponding to the setup data file you specified in advance (see section 11.3,
“Selecting the Power ON Setting”).
If the Instrument Does Not Start Normary When the Power Is Turned On
If the GS610 does not operate as described above when the power switch is turned ON,
turn OFF the power switch and check the following points:
• Check that the power cord is securely connected.
• Check that the voltage supplied from the power outlet is correct. See section 3.3.
• You can initialize the settings of the GS610. See section 11.2.
If the GS610 still fails to power up when the power switch is turned ON after checking
these points, it is probably a malfunction. Contact your nearest YOKOGAWA dealer for
repairs.
Note
If you are turning the power switch ON after turning it OFF, allow at least 5 s before turning it
ON.
For Making Accurate Measurements
At the installation location indicated in section 3.2, allow the instrument to warm up for at
least 60 minutes after the power switch is turned ON before starting the use of the
instrument. Turn the output off during warm-up. After the warm-up time elapses, turn the
output on.