2-12
IM 765501-01E
Auto V/I
(See section 7.7 for the procedure)
This function enables the measurement function to be switched when a limiter is
activated.
For example, when a secondary battery is being charged with constant current, the rising
battery voltage can be measured. When the voltage limiter is activated and the
operation switches to constant voltage charge, the reduction in the charging current can
be measured.
When Generating Voltage and Measuring Current
When the limiter is not activated: The GS610 measures current. The voltage source
level is entered for the source value in the result file
when the storage operation is carried out. (Normal
operation)
When the limiter is activated:
The GS610 measures voltage. The current limiter
value is entered for the source value in the result file
when the storage operation is carried out.
When Generating Current and Measuring Voltage
When the limiter is not activated: The GS610 measures voltage. The current source
level is entered for the source value in the result file
when the storage operation is carried out. (Normal
operation)
When the limiter is activated:
The GS610 measures current. The voltage limiter
value is entered for the source value in the result file
when the storage operation is carried out.
2.4 Measurement