
37
●
User Calibration Function
It is recommended that the instrument be calibrated periodically.
The instrument can be calibrated.
CAUTION
To avoid electrical shock
• Only authorized engineers are allowed to calibrate the instrument
using dedicated facilities.
• Connect the calibrator to the instrument with the calibrator’s testing leads.
• Before carrying out calibration, read the instruction manual of the calibrator.
• Temporarily remove the testing leads from the instrument before switching
measurement mode (function).
<Conditions of calibration>
Calibrator: With accuracy higher than of this instrument
Ambient Environment:
Temperature: 23
±
3ºC
Humidity: 55%RH or less
Leave the instrument for 30 minutes under above conditions
before carrying out calibration.
After reference valve of Calibrator stabilizes, Press the key to confirm for
calibration valve.
Содержание 734 01
Страница 2: ......