Calibrating Using a Shunt Resistor
Shunt calibration can be performed by combining with the YOKOGAWA's strain module
(Model 701271 (STRAIN_DSUB) that supports shunt calibration.
Connection procedures for the shunt resistor
The following two connection methods are available. In normal cases, connect the
resistor in correcting the gain on the negative side.
7
9
8
3
2
5
6
4
1
Bridge-
Input+
Input-
Sense+
Shuntcal
ON/OFF
Sense-
S
Shuntcal-
S
Shuntcal-
R
R
R
7 8 9
7 8 9
Bridge
voltage
Strain
gauge
Example for single-gauge method
Shunt resistor
(negative side)
Shunt resistor
(positive side)
Shunt resistant
connection terminal
Shunt resistant
connection terminal
Shunt resistor
Shunt resistor
When correcting the gain on
the negative side (normal)
When correcting the gain on
the positive side