HF Compensation

HF needs to be adjusted when the probe is connected to the scope input the 

first time.

HF adjustment is performed by connecting the probe to a rectangular wave 

generator with first rise time. Adjust the trimmer for optimum wave response.

 

Optimum

DC Adjustment

In order to provide highest accuracy this probes dividing ratio is factory adjusted 

using 500 VDC source and measuring device with a precision input impedance 

of 1 MΩ ± 1%.

 

LF Compensation 

DC

HF Compensation

Maintenance

Cleaning

When cleaning the exterior of the probe, clean it using a soft cloth rag moistened 

with either water or isopropyl alcohol. In this case, dry the probe completely 

before starting the measurement.

Changing the Contact Tip

To change the contact tip, grip the contact tip firmly with pliers and carefully 

pull it straight out of its contact socket along the axis of the probe. Do not grip 

the white plastic insulator or the housing with pliers, because this will damage 

the probe. If the contact tip is removed, the new contact tip can be inserted 

with pliers into the socket, along the axis of the probe. To insert the contact 

tip completely into the housing, press  the contact tip against a hard surface 

carefully.

 

Use pliers to grip and pull the contact 

tip carefully out of its contact socket.

Do not grip the white plastic insulator 

or the housing with pliers.

Содержание 701944

Страница 1: ...User s Manual Model 701944 701945 100 1 Probe IM 701944 01E 7th Edition ...

Страница 2: ... us to serve you best by completing the user registration form accessible from our website https tmi yokogawa com support Contact Us If you want to resolve a technical support issue or need to contact YOKOGAWA please fill out the inquiry form on our website https tmi yokogawa com contact PIM 103 06E A5 ...

Страница 3: ...ut prior notice as a result of continuing improvements to the product s performance and functionality The figures given in this manual may differ from those that actually appear on your product Every effort has been made in the preparation of this manual to ensure the accuracy of its contents However should you have any questions or find any errors please contact your nearest YOKOGAWA dealer Copyi...

Страница 4: ...categories II defined in IEC 61010 031 If this probe is operated in a manner not specified in this manual this may cause the protection capability of this product to lessen YOKOGAWA shall not be held responsible for defects arising from negligence of such warning and caution and also shall not guarantee the product in such case Before using this probe thoroughly read the instruction manual for mea...

Страница 5: ...imum input voltage to the probe When the oscilloscope s input coupling is AC a DC voltage is applied to the oscilloscope s input at the same electric potential as the probe s input Make sure not to exceed the oscilloscope s maximum input voltage French AVERTISSEMENT But du produit Le produit est utilisé en association avec un oscilloscope pour observer et mesurer des signaux électriques Ne l utili...

Страница 6: ...d entrée maximale sur la sonde Lorsque le couplage d entrée de l oscilloscope est AC une tension DC est appliquée à l entrée de l oscilloscope au même potentiel électrique que l entrée de la sonde Assurez vous de ne pas dépasser la tension d entrée maximale de l oscilloscope Regulations and Sales in Various Countries and Regions Waste Electrical and Electronic Equipment WEEE EU WEEE Directive vali...

Страница 7: ...ptional Accessories Standard Accessories Optional Accessories Probe unit 701944 or 701945 Flexible adapter Spring tip 4mm plug ground lead Rigid tip Safety alligator clip Pincher tip Insulation cap Protection cap Ground lead Color marker 4 colors x 3 Adjustment tool Manuals Attached to the probe unit Replaceable Parts Set Two sets of replaceable parts are available Set name Accessories Basic Part ...

Страница 8: ...lisez le fil de terre uniquement pour les connexions de mise à la terre N utilisez pas d accessoires autres que ceux spécifiés Adjustment Tool Ground Lead length 22 cm 4 mm Plug Ground Lead length 22 cm option Pincher Tip Flexible Adapter option Spring Tip Ø 0 8mm Rigid Tip Ø 0 8mm Insulating Cap Protection Cap Safety Alligator Clip option Probe Unit Color Marker 4 colors x 3 ...

Страница 9: ... 1 This specification value is obtained when the probe is connected to an oscilloscope having an input resistance of 1 MΩ 1 2 This specification value is obtained when the probe is connected to an oscilloscope having a bandwidth of 500 MHz or greater When the probe is connected to an oscilloscope having a bandwidth of less than 500 MHz depends on the bandwidth of the oscilloscope main unit 3 See t...

Страница 10: ... definitions of measurement categories 2 Pollution degree refers to the degree of adhesion of a solid liquid or gas which deteriorates withstand voltage or surface resistivity Pollution Degree 2 applies to normal indoor atmospheres usually with only non conductive pollution 3 For conformity to environmental regulations and or standards other than EU contact your nearest Yokogawa office PIM 113 01Z...

Страница 11: ...augmente la tension d entrée maximale de la sonde diminue Voir les spécifications pour la tension d entrée appropriée Typical Voltage Derating 0 200 400 600 800 1000 1200 0 1 1 10 100 1000 Frequency MHz Amplitude AC rms V Sinus Input Impedance CAUTION As the frequency of the input signal increases the input impedance of the probe decreases French ATTENTION Lorsque la fréquence du signal d entrée a...

Страница 12: ...Pulse Ratings For pulse measurements make sure to comply with the ratings as shown below 0 1000 2000 3000 4000 5000 6000 7000 8000 0 200 400 600 800 1000 Peak Pulse Voltage V AC rms or DC V RMS vs Peak Pulse Voltage Maximum Pulse Derating 1000 1500 2000 2500 3000 3500 4000 4500 5000 5500 6000 10 100 1000 Duration ms Maximum Peak Pulse Voltage V 10 25 50 Duty cycle ...

Страница 13: ...ensation LF needs to be adjusted when the probe is connected to the scope input the first time LF compensation matches the probes cable capacitance to the oscilloscope input capacitance This matching assures good amplitude accuracy from DC to upper bandwidth limit frequencies A poorly compensated probe clearly influences the overall system performance probe scope and introduces measurement errors ...

Страница 14: ...th rag moistened with either water or isopropyl alcohol In this case dry the probe completely before starting the measurement Changing the Contact Tip To change the contact tip grip the contact tip firmly with pliers and carefully pull it straight out of its contact socket along the axis of the probe Do not grip the white plastic insulator or the housing with pliers because this will damage the pr...

Отзывы: