Specifications
Product Specifications
Accuracy of 23 ±3 °C, 30 minutes after turning ON the power.
Model
PBC100 (701928)
PBC050 (701929)
Frequency band*
DC to 100 MHz (–3 dB)
(See the typical characteristics
shown on page 11)
DC to 50 MHz (–3 dB)
(See the typical characteristics
shown on page 11)
Rise time*
3.5 ns or less
7.0 ns or less
Propagation delay
13 ns (Typical)
13 ns (Typical)
Maximum continuous
input
30 Arms (AC and DC components)
(See pages 11 and 12, Derating according to frequency)
Maximum peak current
50 Apeak, non-continuous
Output voltage rate*
0.1 V/A
Amplitude accuracy*
0 to 30 Arms: ±1.0% of reading ±1 mV
30 Arms to 50 Apeak: ±2.0% of reading
(DC, 45 to 66 Hz)
Noise*
Equivalent to 2.5 mArms or less
(with 20 MHz bandwidth measuring instrument)
Insertion impedance
(See typical characteristics on page 12)
Compatible interface
YOKOGAWA probe interface
Temperature coefficient
for sensitivity*
±2% (with 50 Hz, 30 Arms input in a range of 0 to 40 °C)
Maximum rated power
5.5 VA (within maximum input) 5.9 VA (within maximum input)
Rated supply voltage
±12.3 V ±0.3 V
(supplied from the YOKOGAWA probe interface)
Operating temperature
and humidity
0 to 40 °C 80% RH or less (no condensation)
Storage temperature
and humidity
–10 to 50 °C, 80% RH or less (no condensation)
Operating location
Indoor use, altitude up to 2000 m
Effect of external
magnetic fields
Equivalent to a maximum of 5 mA
(in a DC or 60 Hz, 400 A/m magnetic field)
Measurable conductors
Insulated conductors
Diameter of measurable
conductors
φ 5 mm or less
Guaranteed accuracy
period
1 year (up to 10000 open and close operations)
Cable lengths
Approx. 1.5 m (full length: approx. 1.755 m)
External dimensions
Sensor:
approx. 175 (W) x 18 (H) x 40 (D) mm
Probe interface: approx. 80 (W) x 30 (H) x 28 (D) mm
Weight
Approx. 190 g
Accessories
User’s manual, carrying case
* In conjunction with a waveform measuring instrument having an input impedance of 1
MΩ ±1%