Operation Procedure

 

The  Meter has been  calibrated before  leaving  the factory with  built

-

in curve. The  user can directly determine  the 

turbidity value of the sample according to the following methods.

 

1. Press START to turn it on.

 

2.  Pour  the  reference  solution  into  a  random  colorimetric  bottle,  no  more  than  two

-

thirds  of  the  volume  of  the 

colorimetric  bottle.  (Wipe  off  the  residual  liquid  on  the  surface  of  the  bottle  with  a  mirror  cleaning  paper.)  Then 

cover the lid of the colorimetric bottle, screw it down and put it into the sample tank. Pls note that there is a clip in 

the middle. The measurement can only be started after the clip is fixed well.

 

3. Press "UP" button to adjust zero, then take out the colorimetric bottle.

 

4. According to the above method, pour the sample solution into the colorimetric bottle and put it into the sample 

tank.

 

5. The meter displays the turbidity value (NTU) of the sample, Press

ENTER

to store the value in the meter.

 

6. Long press

ESC

key to turn off the machine.

 

Operation Panel

 

ESC

Return, cancel, test stop key

 

ENTER

Enter, function, menu selection key

 

】【

Up and down key

 

SET

Menu view and settings key

 

START

Boot key

 

 

Operation Menu

 

1. Check curve

 

2. Select curve

 

a. A curve is built into the meter before leaving factory. when you establish another curves for your own needs, you 
can select the curve needed for measurement here.

 

b. Press

SET

, then press Select Curve, open curves list, press

/

  to select the desired curve name, support 

the self

-

built fitting curve, press

ENTER

to confirm and open.

 

c. Press

ENTER

again to confirm the selection and automatically return to the measuring interface.

 

3. New curve

 

a. When the built

-

in self built fitting curve of the meter can not meet the measurement requirements, press

SET

key and select new curve.

 

b. The meter can create new single point curve, sectional curve, self

-

built fitting curve and single point correction 

curve. User can choose the corresponding operation according to needs.

 

c. The newly created single point curve does not support for user.

 

d. New preset segmented curves do not support for user.

 

e. Self built fitting curve refers to the function of selecting new single point curve or segmented curve and making 