18
Note: (a) The meter has been calibrated before leaving factory, so generally users
can use it directly.
(b) The standard solution method is aimed at the long time use of
conductance electrode, which leads to inaccuracy of conductance constant.
The new conductance electrode has been calibrated before being released.
Its constant is marked on the electrode and input constant before use.
(c)The electrode constant input method ,first determines the type of electrode
constant. Then, when the input constant coefficient is constant, for example,
the constant is 10.5, select the 10 electrode type first, then set the constant
1.05, that is: 10.5=10*1.05.
6.3. Conductivity solution Test:
6.3.1. Wash the conductivity electrode and throw off the water on it, then insert it into the
solution with temperature electrode, still it after stirring, take the reading after
“
”
appears, which is the conductivity values of this solution.
6.3.2 Short-time press
key can display the value of the TDS, salinity or resistivity
value which against the conductivity value.
6.4. Important Statement:
6.4.1. This meter build-in below two kinds of calibration solution series, please set up in the
parameters P4;
(a)
(Europe & U.S.A. series) — 84μS/cm, 1413μS/cm, 12.88 mS/cm and 111.9
mS/cm
(b)
(China series) — 146.6μS/cm, 1408μS/cm, 12.85mS/cm and 111.3 mS/cm