2.5 Environmental Specifications Temperature
Temperature
Operating.........................................................+50~+104
°
F / +10~ +40
°
C
Storage/Transportation.........................................-4~+140
°
F / -20~+60
°
C
Humidity
Operating.................................................................15~95%, noncondensing
Storage/Transportation............................................10~95%, noncondensing
Atmosphere Pressure
Operating.......................................................................................70~106kpa
Storage/Transportation...............................................................50~107.4kpa
2.6 Display
Display Color ................................................................................
YM201: 0.96",Yellow&Blue
YM301: 1.3",Blue
Display content: SpO2%, Pulse Rate, PI%, Bar Graph Battery Indicator,Pulse Wave
Notes:
1) The claim for oxygen saturation accuracy should be supported by clinical studies
covering the entire claimed range,The fraction of inspired oxygen (FiO2) delivered to
test subjects is varied to achieve a series of targeted steady-state saturation periods
over the specified SpO2 accuracy range (e.g. 70 % to 100 % ),then the SpO2 accuracy
is calculated by comparing SpO2 readings of the pulse oximeter to the values of
SpO2 determined with a Co- Oximeter.