V3.0 OM- VISION Pro valid from 01
st
February 2020
17
Note
The same batch quality controller need to have only one control information.
L-J:
L-J chart takes the test time as the horizontal axis, the test result as the
longitudinal shaft. It shows the controller result tendency in the specified
time period. The value must be between the "Mean-3SD" and the
"Mean+3SD".
Fig. 4-2-3-3 L-J Interface
Click L-J chart, the interface shows the using instrument code, controller
level, controller batch number, control date and L-J chart, as shown in the
following:
Controller time is selected by month, so that L-J chart displays all the
controller information in the selected month.
Controller L- J chart shows all the test results for the controller in the
same batch, and the average and standard deviation correspondingly.
"Curve":
Select a QC result and click “Curve” to view the ESR dynamic
curve.
“Print”:
Select a result and click “Print” to print the result.
“Export”:
Select a result and click “Export” to export the result to a U disk.
4.2.3.4 Quality Control test application
Connect the instrument power cord. For the limited-activated instrument.