V2.3.0 OM-UNN valid from 15
th
April 2015
12
Yellow (Cleaning solution)
Orange (Sanitizing solution)
Red (Waste tank)
Fig. 2-2 Solution Tanks
Note
During Cycle, all 3 solutions (Washing Buffer A, Washing Buffer B and Cleaning solution are used.
For Sanitizing procedure,
-
Cleaning solution is replaced by the Sanitation solution.
-
Both Cleaning and Sanitizing Solution share the same position.
Caution
Handle the solution tanks gently to avoid damage to the weight sensors located at the bottom of the
compartment. Avoid dropping the bottles on the weight sensors or overload the sensors.
Important
All solution tanks must be positioned in its right position to avoid cross-contamination.
2.3.1
Washing Solution
Washing Buffer A tank is color coded with a GREEN cap and holds up to 2 liters in capacity.
Washing Buffer B tank is color coded with a BLUE cap and holds up to 2 liters in capacity.
All Washing Buffers are delivered as concentrates and they should be diluted by DI water and loaded in its
corresponding tanks before use.
Place the tanks in their corresponding slots.
Connect the green labeled tube to washing buffer A tank.
Connect the blue labeled tube to washing buffer B tank.