Cross threading:
Ensure thread is correctly engaged before screwing tight. If not correctly engaged
the thread will be cross threaded and damaged.
Connectors:
Connector wires are small and can be damaged. Always pull the connector apart by
holding the connector not the wires.
The battery voltage may be checked in the software by pressing the Check Logger button. The
nominal battery voltage is 9.0 volts however the actual voltage will range from 9.6volts down to the
cut off of 7.5 volts.
The battery chemistry is alkaline which have some specific qualities
When not using the probe disconnect the battery. 6 months life when connected
Do not leave discharged batteries installed in logger for extended periods
Getting to the battery pack:
1.
Remove cable from top connector.
2.
CAREFULLY UNSCREW
top of probe exposing connectors for battery, main power for probe
and communication. Internal wires are small and will not support the weight of the probe
or top housing.
3.
Track wires from battery and carefully disconnect at the connector. Pull the connector
apart when holding the connector not the wires.
4.
At this point the battery pack can be pulled out and swapped
5.
Install Battery:
1.
Collect/bundle wires from inside probe cavity together.
2.
Push battery pack into cavity guiding the probe wires to one side of the battery pack.
3.
Push connector from battery pack down one side between the battery and the cavity wall
to get it out of the way.
4.
Push top cover into position and screw round knurled nut down
5.
It is a good idea to rotate counter clockwise till the thread gets to the start of the thread
(can be a small click felt or heard). Rotate in the clockwise direction to tighten thread.