Using the OSICS SWT APC Module
162
OSICS User Manual
d. Turn the rotary knob clockwise to put the cursor under
ENTER
and press the
knob to validate the new wavelength value.
The
Unit Setup
menu is automatically displayed and the wavelength/frequency value
of the module is now modified (according to the module limits).
12.7.2 Step-adjusting the Wavelength/Frequency Value of the Module
Subject
The following procedure explains how to modify the wavelength/frequency of the
module step by step using the increment defined in the
Set Step
sub-menu.
You can define a valid step value within the range 00.00–50.00 nm. If you use two non-
adjacent T100 or ECL, check that the step is high enough.
The starting wavelength is the value originally defined in the
Set
λ
/f
Specifying the Optical Emission Wavelength/Frequency, p. 161
Procedure
1.
Access the SWT APC
Unit Setup
menu (see section
Accessing an OSICS Module Setup
).
2.
Expand the
Step
menu and enter the
Set Step
sub-menu.
The
Set Step
sub-menu appears and displays the current value of the wavelength/
frequency step in nm or GHz .
Figure 119: SWT APC Unit Setup – Set Step sub-menu
3.
Modify the step value as follows:
a. Turn the rotary knob to put the cursor under the digit you want to modify and
press the knob to highlight it.
b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and
press the knob to validate the selected digit.
c. Perform steps a. and b. again with the other digits to modify.
d. Turn the rotary knob clockwise to put the cursor under
ENTER
and press the
knob to validate the new wavelength value.
The
Unit Setup
menu is automatically displayed.
4.
To increase or decrease the wavelength/frequency value defined in the
Set
λ
/f
menu
Specifying the Optical Emission Wavelength/Frequency, p. 161
) step by
step, expand the
Step
menu and put the cursor before the
Step +
or
Step -
sub-
menus and press the knob to step-adjust the value.
While the module is updating the wavelength/frequency, the message
Moving…
appears along with the new wavelength/frequency value after step-change.
5.
Repeat step 4. until the wanted wavelength/frequency is reached.
Main
EXIT
1
2
3
4
5
6
7
8
Unit Setup - Set Step (nm)
ESCAPE
▲
Step:
01.00 nm
ENTER
Modify the value.
Содержание OSICS
Страница 2: ......
Страница 6: ...About This Manual 6 OSICS User Manual...
Страница 28: ...Installing and Connecting the OSICS Multifunction Platform 28 OSICS User Manual...
Страница 56: ...Using the OSICS ATN HP Module 56 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 64: ...Using the OSICS BKR Module 64 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 98: ...Using the OSICS DFB Module 98 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 136: ...Using the OSICS SLD Module 136 OSICS User Manual BKR TLS 50 ATN DFB ECL T100 SLD SWT TLS AG SWT APC MAIN...
Страница 248: ...Certification and Compliance 248 OSICS User Manual...