Internal calibration
–
Turn on the device.
Note:
If foreign language text appears, please see page 177 to
find out how to change the language.
N-Tester
©2013
AP 2.02 30002366
The following is shown in the display:
▪
© 2013
: year of the validity of the software
▪
AP 2. 02:
Version of the software
▪
3000 2366 :
Serial number of the device
Pinch empty
sample head for
about 2s.
An internal calibration is necessary each time the device is
switched on. No leaf must be inserted.
The calibration was successful when a single beep can be
heard.
Calibration
is faulty.
Turn off the
device.
If three consecutive beeps can be heard, the calibration
was not successful. You may have opened the measuring
head too early or you have not pressed them together
completely.
–
Switch the device off and back on and repeat the
internal calibration.
Action
After successful calibration you can choose the next action.
Action
Measurement
Settings
–
Select
Measurement.
–
Press .