OXYGEN CONCENTRATOR USER MANUAL
23
NOTE-1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE-2
These guidelines may not apply in all situations. Electromagnetic is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio
(
cellular/cordless
)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the CP101 OXYGEN CONCENTRATOR is used exceeds the
applicable RF compliance level above, the CP101 OXYGEN CONCENTRATOR should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the CP101 OXYGEN CONCENTRATOR.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
ENVIRONMENT PROTECTION
1
、
Treatment of waste and residue
Do not throw away Nasal Cannula
、
humidification bottle and other
component
,
can take them to the professional medical center nearby
.
2、
When the machine is scrapped, please contact your local supplier or
manufacturer.
3
、
The disposal of wastes and residues should be in accordance with the
relevant state laws and regulations.