Indicator light status description
Symbol
Set
E2H
E2P
Erd
Status
ON
OFF
ON
OFF
FLASH
ON
OFF
FLASH
ON
OFF
ON
ON
ON
ON
OFF
Meaning
Parameter setting
Status of temperature measuring and controlling
Compressor high temperature limit alarm (compressor is running)
Capillary heating starts
Capillary heating is closed
High temperature alarm of secondary system condenser
High temperature protection of secondary system condenser
Door open alarm
Grid power supply anomaly
Normal power supply
Parameters
4
Temperature parameter setting
Compressor high temperature limit alarm (compressor is not running)
Compressor high temp. limit alarm reached, unit shut compressor off to cool
Compressor low temperature limit alarm (compressor is running)
Compressor low temperature limit alarm (compressor is not running)
Compressor low temp. limit alarm reached, unit shut compressor off to thaw