Thank you for purchasing the “VS4 Wi

-

Fi Network Bridge”. This 

device enables remote communication with a PC or mobile 

device and allows remote equipment operation, temperature 

monitoring and recording (Data Logger function). 

 

This device works exclusively with Yamato VS4 and VS6 (on 

VS4 mode) controller available on mechanical convection 

equipment models 

DVS/DKN/ADP/DP/FO/IC

 

 

Содержание WBV4A

Страница 1: ...i Network Bridge This device enables remote communication with a PC or mobile device and allows remote equipment operation temperature monitoring and recording Data Logger function This device works exclusively with Yamato VS4 and VS6 on VS4 mode controller available on mechanical convection equipment models DVS DKN ADP DP FO IC ...

Страница 2: ...the FCC identification number is not visible when the module is installed inside another device then the outside of the device into which the module is installed must also display a label referring to the enclosed module This exterior label can use wording such as the following Contains FCC ID 2AHMR ESP07S any similar wording that expresses the same meaning may be used This equipment complies with...

Страница 3: ...and 14 11 4 3 6 Bridge Locate commands ID on ID off 11 4 3 7 Read VS 4 Register command RABC models A and B 11 4 3 8 Write VS 4 register command WABCnnnnn models A and B 11 4 3 9 Saving an Oven program into a file on the Wi Fi Bridge command PnS 12 4 3 10 Loading a program file into Oven Command PnL 12 4 3 11 Starting Fixed Temperature operation remotely Command SFT ttt models A and B 12 4 3 12 St...

Страница 4: ...ure 12 JavaScript editor page 15 Figure 13 working with files 16 Figure 14 Choose file button 17 Figure 15 OS file browser 17 Figure 16 the Upload button 18 Figure 17 verifying the upload 18 Figure 19 Firmware update page 19 Figure 20 Firmware file selection 19 Figure 21 Firmware uploading 19 Figure 22 Remote Operations page 20 Figure 23 Fixed temperature operation 21 figure 24 Program Temperature...

Страница 5: ...any web browser remotely Ex Temperature logs for any past date The software API Application Programming Interface adds convenient functions like Program Upload download from a PC in browser Program editing save back Real Time Status monitoring data logging etc Because it is built on top of HTTP Web socket server standards adding even more functions to the final application is easy for an HTML Java...

Страница 6: ...e receptacle should be inserted on vacant positions of Terminal Blocks TM1 labeled 1 and 2 The two wire communication harness should be connected to VS 4 CN3 pins 3 and 4 If you received a unit with a shielded cable on a 4 pin header block plug this connector on VS4 connector CN3 so that the coaxial cable sits above CN 3 pins 3 and 4 not pins 1 and 2 If your device has a two wire connector the Bla...

Страница 7: ... this module to the cable coming from the Power supply module and attach it to the side or top of the oven This module has a magnetic base that should be used to keep the module attached to the oven s metal chassis Best performance can be achieved by placing the Wi Fi module on top or on the right side face near the top and close to the oven door with the antenna pointed up Avoid spots that get un...

Страница 8: ...t for the first time To configure the Wi Fi Bridge to connect to an Access point the following procedure should be used If there is no Access Point name pre programmed on the Wi Fi Bridge or if it cannot connect to a previously configured Access Point the bridge enters Wi Fi Setup mode It starts its own Access point and activates its bright Red LED light The installer should use a PC or cell phone...

Страница 9: ... will always try to use that same IP even after Power off Before the Main Menu screen is displayed an authentication pop up appears It asks for Username and Password information For now the Use name is fixed at WBVuser and password is Yamato After entering these credentials the Main Menu screen should be displayed There you can easily go to the local File Editor tool see temperature log charts upd...

Страница 10: ...VS 4 6 Wi Fi Bridge 2021 03 09 Page 9 FIGURE 7 WI FI BRIDGE MAIN MENU MODEL A FIGURE 8 WI FI BRIDGE MAIN MENU MODELS B AND C ...

Страница 11: ...l A for 30 days After that the unit will assume the identity of Model C unless the user activates it with a License For model A or B The customer should have received a License file corresponding to the purchased model by e mail If that has not happened or customer wishes to upgrade the unit to a model B or A contact Yamato USA To activate the License file on a WBV4 unit use the File Editor utilit...

Страница 12: ... Wi Fi settings and cause a re start Upon re booting the Bridge will start its internal Acess point and the user will need to re configure the local Wi Fi AP name and password just like on the initial installation procedure 4 3 5 Bridge information command 14 Enter 14 at the command line to see information on your Wi Fi Bridge This may be useful on a technical support call The same information is ...

Страница 13: ... line enter SFT nnn where nnn is the three digit target Temperature The oven will change state from Stop to Run and the heating element will be turned on The oven will heat up until the Target temperature is reached It will then keep the chamber at that set temperature until the Stop command is issued This function is also available on the Remote Operations page graphical interface 4 3 12 Starting...

Страница 14: ...window as follows 5 1 Log nn command The Log command can be issued without parameters in which case it returns the Log operation state and the current log interval When used with the nn parameter it sets the number of minutes between data point captures If nn is set to 00 logging will not occur disabled NOTE Whenever the user changes the Log interval or enables disables Logging the Wi Fi Bridge wi...

Страница 15: ...e sample rate is set to 4 minutes and the oven only operates 8 hours per day by default capture only occurs if the oven is ON and temperature inside the chamber is 25 Celsius In this case the bridge can keep about 2 years of data in its Flash memory before it fills up By default the Home page of WBV4 model A displays Today s temperature chart as taken from the log file corresponding to Today s dat...

Страница 16: ...issued on Bridge s main web page FIGURE 12 JAVASCRIPT EDITOR PAGE 6 1 Standard file names and descriptions The Wi Fi bridge has a simple built in File System that stores files necessary for the bridge configuration its Web interface and other programs for communicating with an oven For security reasons only txt files are displayed readme txt almost same as the Help command but may contain revision...

Страница 17: ...ll cause the browser to download the file into the PC or phone that started that operation On this example Program3 file is being edited NOTE Deleting a file cannot be undone FIGURE 13 WORKING WITH FILES NOTE Another way to download files from the Bridge to a computer without using the editor utility is to point your browser directly to the file you want Ex to download file Program2 txt just navig...

Страница 18: ...ication and then upload it to the Wi Fi bridge To Upload a file click on the Choose File button on the upper left corner of the page FIGURE 14 CHOOSE FILE BUTTON An Operating system Pop up dialog will open and the user can navigate the PC file system and find the correct file Select the file to upload and click Open FIGURE 15 OS FILE BROWSER ...

Страница 19: ...he selected file name shows up on the text box next to the Upload button Click the Upload button to execute the opetation FIGURE 16 THE UPLOAD BUTTON Select the just upload file on the left side on the editor window to load it on the editor and verify that the correct file was uploaded FIGURE 17 VERIFYING THE UPLOAD ...

Страница 20: ... update c Click Firmware update button on Wi Fi Bridge main menu d Select the new bin file by clicking Chose File button An OS pop up window help you with that e Back to the Update firmware web form the Update button f You can track the file upload progress on the bottom of the browser window g Once the firmware file has been uploaded the Bridge will validate its content MD5 hash and proceed to se...

Страница 21: ...ote Operations page is displayed Currently two modes of operation can be initiated monitored Fixed Temperature mode and Program mode A dynamically updated status frame is always displayed on the lower part of the browser window It updates itself every 10 seconds Once the operations are started the oven is in full control and the web browser window can be closed without interfering with the on goin...

Страница 22: ...he Wi Fi Bridge is able to load and save back all three operating programs stored in the connected oven Use the Command Line window to load existing programs from an Oven into the Wi Fi bridge PnL where n is the program number Likewise the command PnS will save a program residing on the Wi Fi bridge into the connected oven The Save back to oven operation is done automatically when the remote opera...

Страница 23: ... button For simplification purposes the rise time is assumed to be three degrees min Fall time is assumed at two degree min The chart below is real life approximation the Program shown on the bottom of this page Note the repeat loop after Step 3 which executes twice The total time to run the program is considerably longer once Rise Fall times are factored in Program1 txt shows total time of 150 mi...

Страница 24: ...ther PCs or cell phones on same network To design these Dashboards or Flows Node red s browser based front end provides an extensive library of graphical elements representing logical functions of various complexities and the programmer just have to interconnect them to implement the desired behavior On the example below a simple Oven management interface allows operators to start an oven remotely...

Страница 25: ...e um R1 3 24 Dashboard for Fixed Temperature operation Dashboard for Program mode operation FIGURE 28 FIXED TEMP DASHBOARD Status dashboard after Start Prog mode FIGURE 29 STATUS DASHBOARD FIGURE 27 PROGRAM MODE DASHBOARD ...

Страница 26: ...al or consequential damages of any character arising as a result of the use of this device and software or out of the use or inability to use the Software including but not limited to damages for loss of goodwill work stoppage computer failure or malfunction or any and all other commercial damages or losses even if such Contributor has been advised of the possibility of such damages Note Contents ...