 VENTED BALANCE ENCLOSURE

 

Model

VBE204/214/306/316/408/418/600/610 

Instruction Manual 

– December 2019 –

Thank you for purchasing the Vented Balance

Enclosure, VBE204/214/306/316/408/418/600/610,

of Yamato Scientific.

To use this unit properly, read this "Instruction

Manual" thoroughly before using the unit.

Keep this instruction manual around this unit for

reference at any time.

 WARNING: 

Carefully read and thoroughly understand the 

important warning items described in this 

manual before using this unit. 

Содержание VBE204

Страница 1: ...losure VBE204 214 306 316 408 418 600 610 of Yamato Scientific To use this unit properly read this Instruction Manual thoroughly before using the unit Keep this instruction manual around this unit for reference at any time WARNING Carefully read and thoroughly understand the important warning items described in this manual before using this unit ...

Страница 2: ...tents General Information 2 Set Up Procedures 4 Exhaust Connection Filter Change Out 5 General Maintenance Cleaning and Sterilizing 6 Models Specifications 7 Replacement Parts 8 After Service and Warranty 9 ...

Страница 3: ... assembly includes a backward curved impeller panel HEPA Filter speed control on off switch and house exhaust connection During normal operation the backward curved impeller maintains a face velocity into the enclosure which prevent particulates from escaping The impeller pulls air into the enclosure through the back baffle and then through the HEPA filter Any particulates that are generated insid...

Страница 4: ...3 General Information ...

Страница 5: ... 7 Turn the main power switch on The blower speed control has an off position If the speed control is in the off position the blower will remain off 8 Place an anemometer air flow meter on the front face of the enclosure The meter should be 50mm from the front face Adjust the blower speed control so the face velocity is 70 FPM 0 355 m sec 9 Plug in the air flow alarm into a 110 220V socket 10 Cali...

Страница 6: ...Use a vinyl tape and tape the front and back sides of the top plenum panel to the top of the VBE 5 Remove the air flow monitor from the side of the filter housing 6 Lift off the top HEPA filter fan assembly and place on bench top 7 Place bag P N 900 VBE BAG over the top of the filter housing and tape to the side of the filter housing 8 Using the bag pull the filter out of the filter housing and up...

Страница 7: ... This applies to all plastics used in these products Examples of these products include but are not limited to acetone ether gasoline lacquer thinner methyl ethyl ketone methylene chloride and toluene Polishes While the above cleaning solutions have some polishing capabilities they will not remove scratches from plastic This can only be done with automotive type waxes or the finer grades of rubbin...

Страница 8: ...4 30 EC Electromagnetic Capability 89 686 EEC Safety 2014 35 EU Low Voltage EN61326 1 2013 EN55011 2016 Class B EN61000 3 2 2104 EN61000 3 3 2013 EN61000 4 2 2009 EN61000 4 3 2010 EN61000 4 4 2012 Model Inside dimensions Outside dimensions VBE204 VBE214 24 00 w x 23 00 d x 21 50 h 609 w x 584 d x 546mm h 26 00 w x 26 75 d x 41 00 h 660 w x 679 d x 1041mm h VBE306 VBE316 35 25 w x 23 00 d x 21 50 h...

Страница 9: ...EL1682 Impeller 110VAC 530 CFM EL1703 Capacitor 10 uF for 110 VAC impeller ELl683 Capacitor 2uF for 230 VAC impeller 900 BE 24 HEPA HEPA Filter 24 x 18 Rated 99 995 3 µ H14 900 VBE 36 HEPA HEPA Filter 30 x 18 Rated 99 995 3 µ H14 900 VBE BIBO Bag In Bag Out Port 9 900 VBE BIBO BAG Waste Bag for BIBO port 900 VBE BAG S Filter Bag 24 x 20 x 48 900 VBE BAG L Filter Bag 30 x 18 x 48 ...

Страница 10: ...t responsible and will not be held liable for damage and or injury caused using our products Product Return Policy If you are not satisfied with your purchase and wish to make a return contact our customer service to inquire about a Return of Merchandise Authorization Number RMA Merchandise returned without an RMA number will not be accepted and will be returned to the sender Return requests must ...

Страница 11: ...this document forbids as unexpected accidents or breakdown may result Instruction Manual for Vented Balance Enclosures Model VBE204 214 306 316 408 418 600 610 December 2019 Yamato Scientific America Inc 925 Walsh Ave Santa Clara CA 95050 Tel 1 800 292 6286 408 235 7725 http www yamato usa com For customer service Email customerservice yamato usa com For technical support Email technical yamato us...