Hysteresis is set at 3, then the compressor will cycle from 3 to 6 degrees Celsius. To change 
this value, press SET and DOWN keys for 3 seconds to enter programming mode. “Hy” will 
display meaning Hysteresis. Press Set again and the current value will be displayed. Use the 
UP and/or DOWN arrow keys to adjust the value and press SET again to lock it in. Press SET 
and UP to return to Operating Mode. 

 

 

Adjusting other User Adjustable settings: 

-

 

To change between Celsius and Fahrenheit scales, press SET and DOWN keys for 3 seconds 
to enter programming mode. “Hy” will display. Press the UP or DOWN keys several times to 
cycle through the parameters until “CF” is displayed. Press SET to display the current setting 
(either C for Celsius or F for Fahrenheit). Use the UP or DOWN key to choose the desired 
setting. Press SET again to lock it in. Press SET and UP to return to Operating Mode. 

 

To change the Hi or Low Alarm set points, press SET and DOWN keys for 3 seconds to enter 
programming mode. “Hy” will display. Press the UP or DOWN keys several times to cycle 
through the parameters until “ALU” for Alarm Upper (High Temperature Alarm) or ALL for 
Alarm Lower (Low Temperature Alarm) is displayed. Press SET to display the current value. 
Use the UP and/or DOWN key to choose the desired value. Press SET again to lock it in. 
Press SET and UP to return to Operating Mode.

 

 

 

Содержание UCR101

Страница 1: ...gerator of Yamato Scientific To use this unit properly read this Instruction Manual thoroughly before using this unit Keep this instruction manual around this unit for referring at any time WARNING Ca...

Страница 2: ...must be noted on the freight bill and acknowledged by the carrier s agent driver at time of delivery Make sure you get a signed copy Send a written request for an inspection to the carrier B Concealed...

Страница 3: ...y receipt acknowledging the damage This has to happen at the time of delivery or it won t happen at all Keep a copy for your records and send another to the carrier s damage claims department along wi...

Страница 4: ...is cabinet must conform to all National and Local Electrical Codes Consult the cabinet Serial Data plate for voltage cycle phase and amp requirements before making connection VOLTAGE SHOULD NOT VARY M...

Страница 5: ...ecially if the weather is very hot This is normal The motor compressor will cycle normally as soon as the excess heat has been removed TEMPERATURE CONTROL NOTE The control is preset at the factory to...

Страница 6: ...ides of the unit to allow for proper air flow and therefore more even cooling of the product When loading the top shelf avoid blocking the evaporator fan s if installed There should be at least 4 inch...

Страница 7: ...ides the following functions Red LED for visual indication of an alarm condition Beeper for audible indication of an alarm condition A preprogrammed 5 minute delay from the time a set point is attaine...

Страница 8: ...Pon The control keys are now unlocked and you can perform the programming functions below Adjusting the temperature control set points NOTE The temperature control set points are preset at the factor...

Страница 9: ...isplay Press the UP or DOWN keys several times to cycle through the parameters until CF is displayed Press SET to display the current setting either C for Celsius or F for Fahrenheit Use the UP or DOW...

Страница 10: ...d the proper refrigerator air circulation For the proper air circulation place the products evenly on each shelf Do not push against the refrigerator s rear or side walls Evaporator is blocked by fros...

Страница 11: ...bove 40C 104F Stop using immediately The display shows LL and the compressor does not run The ambient temperature is below 40C 40F Stop using immediately MOISTURE DURING THE SUMMER SEASON The amount o...

Страница 12: ...e the proper tools When cleaning stainless steel surfaces take care to use non abrasive tools Soft cloths and plastic scouring pads will not harm the steel s passive layer Stainless steel pads may be...

Страница 13: ...er Sheen Passive Film Breakdown Corrosion on metals is everywhere We recognize it quickly on iron and steel as unsightly yellow orange rust Such metals are called active because they actively corrode...

Страница 14: ...enheit For Refrigerator Press and hold DOWN SET until Hy appears on display Now release the buttons Press and hold DOWN SET again until Pr2 appears The Pr2 will appear briefly and disappear when you l...

Страница 15: ...to Hy Press UP or DOWN until CF is displayed press SET Change C to F and press SET This changes Celsius to Fahrenheit Then press UP or DOWN until LS is displayed press SET Low set point limit Change t...

Страница 16: ...s displayed press SET Upper set point limit Change the value to 60 for factory setting and press SET Then press UP or DOWN until ALU is displayed press SET Upper alarm setting Change the value to 50 f...

Страница 17: ...Hy is displayed press SET Differential Change the value to 4 for factory setting and press SET Do not push any other buttons and let the display return back to normal operation Press and hold SET unti...

Страница 18: ...Conditions This warranty shall have force and effect only if all items are used with proper circuits voltages and frequencies and the operation thereof is in accordance with instructions furnished by...

Страница 19: ...to a 25 restocking fee Shipping and handling cost are non refundable All retrofitted customized and special order item sales are final and non returnable In Case of Request for Repair If the failure...

Отзывы: