YCO-1052-03-00

PB-200

                      Technical Manual

5

III Troubleshooting

Problem

               

 

 

The scale will not power on.

The dis play con tin ues to display 
four eights af ter the scale is 
turned on.

“Err” appears upon power up.

“ErrL” appears upon power up.

The displayed weight fl uc tu ates 
ex ces sive ly.

Possible Cause 

 

 

              

Batteries dead or incorrectly 
installed.
AC adapter loose or mal func -
tion ing.
CPU board malfunction.

Vibration or an unstable po-
 si tion or platform.

Platform weight too high.

Platform weight too low.

Material between the platform 
and the scale body.
Weighing assembly in contact 
with wires or case.
Load cell connector loose or 
damaged.
Power board malfunction.
CPU board malfunction.

Solution 

               

 

Test batteries.  Reinstall if 
needed.
Reseat adapter.  Replace if 
needed.
Replace CPU board.

Remove source of vibra-
tion or place scale on stable 
platform.

Clear platform and restart.

Remove platform, clear 
away any debris and re in -
stall platform.

Remove material.

Reposition wires, case or 
platform.
Reconnect or replace the 
load cell connector.
Replace power board.
Replace CPU board.

Содержание PB-200

Страница 1: ...YCO 1052 03 00 YAMATO CORPORATION 1775 S Murray Blvd Colorado Springs CO 80916 USA Tel 719 591 1500 Fax 719 591 1045 YAMATO TECH CORPORATION 112 19425 Langley By Pass Surrey B C V3S 6K1 Canada Tel 604 533 2338 Fax 604 533 0827 ...

Страница 2: ...e Not only will the scale be damaged but you may also be injured 4 Place the item to be weighed in the center of the platform Items placed on the edge of the platform may fall off and cause injury 5 When weighing a heavy large or unbalanced item make sure the item is stable on the platform otherwise an accident may occur 6 When carrying or moving the scale be sure to hold it by the bottom of the b...

Страница 3: ... off including full disable is set in the System Parameter mode see section II A Use of an AC adaptor disables the automatic off feature II A System Parameter Mode II Set up Procedures 1 With the scale on press and hold the key until the display shows Set 2 Release the key The display will now show Unit 3 To disable operation of the scale in any of the four units press the key 4 Pressing the key w...

Страница 4: ... pressing the key until all desired changes have been made 9 Press the key until the display reads Unit 10 Press the key once to proceed to automatic off setting see section II A 2 for details 11 To return to normal operating mode and save any changes press the key twice The display will read End Press the key 12 To exit the system parameter mode without saving any changes press the key II Set up ...

Страница 5: ...e to enter Automatic off adjustment mode 5 Pressing the key will advance the display through 120 180 240 300 and OFF 6 Select the desired setting by pressing the key 7 To return to normal operating mode and save any changes press the key The display will read End Press the key 8 To exit the system parameter mode without saving any changes press the key II Set up Procedures II A System Parameter Mo...

Страница 6: ...ro the counts 3 Select either pounds or grams by pressing the key 4 Place either 10 pounds or 5 kilograms on the scale platform The counts should read 5000 to 9000 for pounds or 5000 to 10 000 for grams 5 Press the key The display will read Cal for about four sec onds 6 The scale should return to normal operating mode and display the calibration weight 7 Remove the weight from the scale platform T...

Страница 7: ...r platform Platform weight too high Platform weight too low Material between the platform and the scale body Weighing assembly in contact with wires or case Load cell connector loose or damaged Power board malfunction CPU board malfunction Solution Test batteries Reinstall if needed Reseat adapter Replace if needed Replace CPU board Remove source of vibra tion or place scale on stable platform Cle...

Страница 8: ...YCO 1052 03 00 PB 200 Technical Manual IV Wiring Diagram 6 ...

Страница 9: ...YCO 1052 03 00 PB 200 Technical Manual V Parts List 7 ...

Страница 10: ...able 1 13 CHA 1585 200120 PCB Display 1 14 CHA 1585 ODS310 AC Adaptor not shown 1 15 CHA 1585 200150 Cable Connecting 1 16 Platform SS 1 17 Screw Hex Head Flat Countersunk M6 x 20 4 18 Plate Upper 1 19 Standoff A D PCB 4 20 Housing A D 1 21 Screw Pan Head M4 x 5 2 22 CHA 1585 200140 PCB A D 1 23 Jack Foot Tare Switch 1 24 Spacer Load Cell 2 25 CHA 1585 200130 Load Cell 1 26 Screw Pan Head M5 x 17 ...

Отзывы: