22
4. Operating procedures
Related Figure between Blower and Temperature/Circulating Air Quantity
(Reference)
The temperature around outlet is depended on the circulating air flow rate of the blower.
Also, this temperature is depended on the clogging of the filter in the blower. Refer to the following
table for each dial value of the blower and the display temperature around inlet/outlet as the
guideline of the work.
Display temperature of outlet (
℃
)
Setting temperature
of inlet
Blower dial value
50
℃
100
℃
150
℃
0
~
4
44
73
97
4.5
44
75
98
5.5
45
76
100
7
45
76
101
8.5
46
78
102
10
47
79
103
The table below shows the relationship between dial values of the blower and average circulating
air amount.
Use this table as a reference for work. When the filter is clogged, air amount for the specific dial
setting may decrease. Clean the filter according to the maintenance method when the reading on
the filter differential pressure meter has become about twice of the ordinary value.
Blower
dial value
Average flow rate
of circulating air
(
m
3
/min
)
0
~
4
0.12
4.5
0.14
5
0.17
5.5
0.21
6
0.24
6.5
0.29
7
0.32
7.5
0.38
8
0.43
8.5
0.49
9
0.53
9.5
0.58
10
0.65
(
11
)
(
0.75
)