31
8. Troubleshooting
When a malfunction is suspected
If any of the symptoms below occurs
Symptom
Check
Turning the ELB to on will
not activate the unit.
●
If the power cord is connected to the power supply securely.
●
If power outage is not occurring.
●
If the standalone overheat prevention device is working.
Temperature does not rise.
●
If the set temperature is below that in the device.
●
If the power supply voltage has declined.
●
If the ambient temperature is not low.
●
If cooling load for inside the bath is not too large.
Temperature fluctuates
during operation.
●
If the set temperature is appropriate.
●
If the power supply voltage has declined.
●
If ambient temperature fluctuates widely.
●
If cooling load for inside the bath is not too large.
Displayed temperature
differs from the
measurement.
●
If the calibration offset setting is not other than “0”. Set it to “0.”
Confirm settings in “Useful functions (calibration offset function)”
in page 23.
If power outage occurs
When the power is applied again after the unit has stopped due to power outage, the unit will
automatically return to the status immediately before the power outage and resumes operation.
Turn the ELB off if you do not want to resume operation by automatic recovery.
◆
If the symptom does not match any of the above, immediately turn the ELB on the main unit off,
pull out the power cord from the power supply and contact your dealer or one of our sales
offices.