4.

 

Operation Method 

Vacuumizing  and  air  inflow   

 

 

 

   
 

 

 

 

Vacuumizing: 

1

 Turn the purge valve CW to its end (close); 

2

 Be sure the pump valve is close, turn on the vacuum pump. 

3

 Turn the pump valve CCW to its end, the exhaust speed is max. 

here.  If  the  samples  are  powder  or  foaming  objects,  please 
slowly turn on or turn on part of exhaust valve. 

Air inflow: 

1

 Turn the pump valve CW to its end (close); 

2

 Turn  the  purge  valve  CCW  to  its  end,  the  inflow  speed  is  max. 

here.  If  the  samples  are  powder  or  foaming  objects,  please 
slowly turn on or turn on part of intake valve. 

3

 The  door  of  chamber  is  unable  to  be  opened  when  the  inside 

pressure resumed to atmosphere pressure. 

 

This  unit  does  not  have  vacuum  breaker  system  and  vacuum 
pump  anti  oil  backstreaming  system.  Please  buy  the  anti  oil 
backstreaming accessories from the vacuum pump agent. 

 

Turn  on  or  off  the  vacuum  pump  when  the  exhaust  valve  is 
close, or else the lube oil of vacuum pump would back stream 
to contaminate chamber and sample. (unless there is vacuum 
breaker) 

 

If  there  is  power  cut,  please  turn  off  the  exhaust  valve  right 
away, or else the lube oil of vacuum pump would back stream 
to contaminate chamber and sample. (unless there is vacuum 
breaker) 

 

The  above  operations  must  be  strictly  carried  on  as  per  the 
procedures.  Our  company  is  not  responsible  for  the  loss  due 
to incorrect operation.

 

 

 
 
 
 

PURGE VALVE

 

Max

 

Min

 

PUMP VALVE

Max

Min

 

 

Before the pressure of suction port of pump recovers to atmosphere pressure, do not turn off 
the  power  switch  of  vacuum  pump.  If  turn  off  the  vacuum  pump  directly,  the  oil  may  back 
stream. 

Notes

 

Содержание DPC Series

Страница 1: ...erica Co Ltd To use this unit properly read this Instruction Manual thoroughly before using this unit Keep this instruction manual around this unit for referring at any time WARNING Carefully read and...

Страница 2: ...Stop Operation 19 Auto Stop Operation 21 Auto Start Operation 23 Program Operation 25 Program Repeat Function 30 Programming Preparation Form 31 Other Functions Calibration 33 Other Functions Lock 34...

Страница 3: ...nger of a problem that may cause a serious accident or even fatality CAUTION If the caution is ignored there is the danger of a problem that may cause injury damage to property or the unit itself Mean...

Страница 4: ...Caution electrical shock Caution scald Caution no load heating Caution not to leak Caution water only Caution deadly poison Prohibit Prohibit generally Prohibit inflammable Prohibit to disassemble Pro...

Страница 5: ...ce it is dangerous and not recommended Do not use the power cord if it is bundled or tangled Do not use the power cord if it is bundled or tangled If it is used in this manner it can overheat and fire...

Страница 6: ...telephone or lightning conductor If not fire disaster or electric shock may be caused M5 The power plug is not accessory Please correctly connect the ground wire as per the power source device 2 Choos...

Страница 7: ...xplosive gas Refer to page 48 13 List of Dangerous Substances Never use this unit in an area where there is flammable or explosive gas This unit is not explosion proof An arc may be generated when the...

Страница 8: ...or malfunction SINCE 1889 DP83C Vacuum Drying Oven SINCE 1889 DP83C Vacuu m Drying Oven 6 Do not make an overload 7 Do not set samples in close formation The withstand load of shelf is 15kg uniform lo...

Страница 9: ...n materials decomposing It is not machine failure It s better to have a high temperature operation before using 11 Handling of power cord Do not entangle the power cord This will cause overheating and...

Страница 10: ...V 25KF flange Stainless steel vacuum rubber hose 1pc Clamp 25KF 1 pc Clamp 40KF 1 pc Center ring with O ring 25KF 1pc Center ring with O ring 40KF 1pc Vacuum quick release reducer joint 25 40KF 1 pc M...

Страница 11: ...tion and Function of Each Part Main Unit DP83C Door handle Door Operation panel Cable hole Caution plate Caster Power switch ELB Pump valve Purge valve Observation window Clasp Vacuum pressure gauge S...

Страница 12: ...scription and Function of Each Part Main Unit DP104C Door handle Door Operation panel Cable hole Fastening handle Power switch ELB Pump valve Purge valve Observation window Vacuum pressure gauge Suppo...

Страница 13: ...TOP Lamp Blink while setting quick auto stop timer or auto stop timer Light while quick auto stop timer or auto stop timer is running AUTO START Lamp Blink while setting auto start timer Light while a...

Страница 14: ...type from 1 to 3 Refer to Page 25 Program Operation PAt Program Pattern Used for choosing program pattern Refer to Page 25 Program Operation End Time Up Displays when the timer operation is completed...

Страница 15: ...erature Refer to Page 33 Other Function oH Overheating Prevention Setting Used for setting temperature for overheating prevention device Refer to Page 17 Setting of Overheating Prevention Device Lock...

Страница 16: ...fixed temperature operation Pressing the TIMER key displays AStp The setting temperature SV can be set by pressing the ENTER key The operation time tim can be set by pressing it again Pressing the ST...

Страница 17: ...again when the POWER switch is pressed again manual reset 2 Calibration offset function This calibration offset function is for calibrating the difference occurred between the required in furnace temp...

Страница 18: ...16 4 Operation Method Operation Mode Function Setting Key and Characters The operation mode and function settings use the operational keys and characters as shown below...

Страница 19: ...the ELB The default value is displayed for about 4 seconds after turning on the power The screen then displays the initial setting The current temperature in furnace operation mode character and sett...

Страница 20: ...mode Press the FIXED TEMP key to display FIX which indicates the fixed temperature operation on the center display screen 3 Set the temperature Press the FIXED TEMP key again The setting temperature...

Страница 21: ...nts of a minute under 99 hours and 59 minutes It can be set in increment of ten minutes over 100 hours The can change the setting time quickly when it is pressed continuously Press them discontinuousl...

Страница 22: ...imer is the time subtracting the elapsed time For example set timer as 1hr30min conduct quick auto stop operation after 20min change the set time to be 2hr set it by TIMER the remaining time is 1hr40m...

Страница 23: ...ess the ENTER key again The measurement temperature display screen displays the character tim which indicates the timer setting The setting temperature display screen displays the current setting time...

Страница 24: ...th the FIXED TEMP and AUTO STOP lamps lighting on Press the START STOP key to terminate the timer operation mode The screen returns to the initial setting screen To correct or check setting Changing t...

Страница 25: ...AUTO START lamp blinks too Set the temperature using the Press the ENTER key again The measurement temperature display screen displays the character tim which indicates the timer setting The setting t...

Страница 26: ...heck setting Changing the setting temperature or time during operation is possible by pressing the TIME key Use the to change the setting value Press the ENTER key respectively after changing the sett...

Страница 27: ...ck the number of steps in a created program and their setting temperature time Refer to page 25 27 Check the temperature rise fall capability of the unit Set the time within the capability above Suppo...

Страница 28: ...tion RT23 humidity 65RH 20 unloaded unit min DP83C Rise time Fall time 200 45 60 150 35 180 100 25 450 50 15 Program creation The program pattern below is explained as an example 1 Program pattern exa...

Страница 29: ...plays End When PrG2 is selected the measurement temperature display screen displays the program pattern PAt1 For the pattern of PrG2 select 1 or 2 using the Press the ENTER key again The measurement t...

Страница 30: ...f the first step is displayed The current setting time is also displayed with blinking Before setting the time check the temperature rise fall capability of unit For example for DP83C about 120 minute...

Страница 31: ...e maximum setting time is 999 hours and 50 minutes The time can be set in increments of a minute under 99 hours and 59 minutes It can be set in increment of ten minutes over 100 hours The can change t...

Страница 32: ...t temperature screen indicates PS 7 in the example because repeat function is used at the seventh step The step number 1 to 7 can be input in the setting temperature display screen Enter the number 1...

Страница 33: ...ion Form Please use this form by making copies Register with PrG1 PrG2 PrG3 PAt1 PAt2 PAt3 No Project Name Date Programmer Program Pattern 30 29 28 27 26 25 24 23 22 21 20 19 18 17 16 15 14 13 12 11 1...

Страница 34: ...Project Name Date Programmer Input Value Setting Temperature Setting Time min Repeat Function To Times Step 1 Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Step 8 Step 9 Step 10 Step 11 Step 12 Step 13 S...

Страница 35: ...the SUBMENU key Select the character cAL which indicates the calibration offset using the and then press the ENTER key Input the difference using the and then press the ENTER key This completes the s...

Страница 36: ...using the and then press the ENTER key The setting temperature screen displays oFF The setting value is locked when it is turned to on using the Press the SUBMENU key again to cancel the lock Select...

Страница 37: ...breaker system and vacuum pump anti oil backstreaming system Please buy the anti oil backstreaming accessories from the vacuum pump agent Turn on or off the vacuum pump when the exhaust valve is clos...

Страница 38: ...ey immediately then turn off the circuit breaker and the main power If this procedure is not followed fire or electrical shock may be caused 3 When open close door Do not get close to the traveling ra...

Страница 39: ...bnormally rise and it will cause trouble Never put sample on the chamber surface Fix the shelf on the bracket then set sample on it 9 Recovering after power failure When power is supplied after a powe...

Страница 40: ...rature Never try to disassemble the unit Caution Wipe off the dirt on resin parts or operation panel with a tightly wrung soft cloth Never clean the unit with gasoline banana oil or detergent otherwis...

Страница 41: ...recycle the reusable parts considering to the environmental protection The feature components of this unit and materials used are listed below Main Component Name Material Main Components of structur...

Страница 42: ...hort circuit detection ALARM lamp lights on Er 02 appears Triac is in short circuit Make a call for service Heater disconnecting detection ALARM lamp lights on Er 03 appears Heater is disconnected Mak...

Страница 43: ...ng load inside the chamber is too large Temperature fluctuates during operation If the set temperature is appropriate If the power supply voltage has declined If ambient temperature fluctuates widely...

Страница 44: ...fices and please fill in your dealer date of purchase and other information and store securely Warranty period is one full year from the date of purchase Repair service for free is available according...

Страница 45: ...gital display Time 1min 99hr 59min and 100hr 999 5hr attached with timing function Time resolution 1 min and 10 min Operation Fixed temp operation program operation quick auto stop operation auto stop...

Страница 46: ...rox 1000kg Accessories Shelf plate Stainless steel SUS304 2pcs 4pcs Instruction manual warranty 1 copy for each 1 The performance is under the condition of AC 220V DP104C is 3 phase 480V power source...

Страница 47: ...CN9 CN8 18 1 1 18 CONT CN2 1 2 3 3 2 2 2 1 4 3 SSR 2 X H 4 3 2 1 1 TM1 1 CN5 5 6 4 4 Symbol Part name Symbol Part name ELB Earth Leakage Breaker CONT Control board TM1 Terminal block PIO Display board...

Страница 48: ...400WX36 Q0 3P 30A 125mA 2 4 6 5 3 1 AC480V 3 4W 50Hz 25A KM1 1 2 3 4 5 6 R11 T11 R12 T12 R S T PE T1 R1 A2 A1 A2 A1 SSR SSR 40A SSR3 CT CTa CTb CTb CTa S11 Control Power R21 S21 Temp Controller S1 Sy...

Страница 49: ...1299002 CT Current transformer URD CTL 6 S 4 YSJ B010509001 SSR Solid state relay G3NB 240B 1 DC5 24V YSJ A011006014 DP83C Replacement parts ELB Earth Leakage Breaker NV L22TZU 24A 2P 30mA YSJ B010410...

Страница 50: ...e Potassium perchlorate Sodium perchlorate Ammonium perchlorate and other perchlorate Potassium peroxide Sodium peroxide Barium peroxide and other inorganic peroxide Potassium nitrate Sodium nitrate A...

Страница 51: ...Specification power supply P 4 P 7 P 43 2 Operation start Starts operation Performs fixed value operation auto stop operation or auto start operation 2 Before operating the unit Installation procedur...

Страница 52: ...xpected accidents or breakdown may result in Note The contents of this document may be changed in future without notice Any books with missing pages or disorderly binding may be replaced Instruction M...

Отзывы: