37
Panel display
Conditions that trigger the error code
Corrective action
<When the optional RS232C is in
use>
This code appears when the
instrument has received a tare
subtraction command, but is not ready
for tare subtraction.
Check the specifications/status
of the instrument. Make the
instrument ready for tare
subtraction and then transmit
the command.
Press the
key to reset the
error and then restart
communication.
<When the optional RS232C is in
use>
This code appears when an
inexecutable command is received.
For example, an addition command is
received when there are no objects
placed on platform of the instrument.
Check the specifications/status
of the instrument. Make the
instrument ready for the
command and then transmit
the command.
Press the
key to reset the
error and then restart
communication.
<When the optional RS232C is in
use>
This code appears when an undefined
ZS99 command is received.
Check the data transmitted by
the connected equipment.
Press the
key to reset the
error and then restart
communication.
<When the optional RS232C is in
use>
This code appears when the
transmitted data length is too long.
Check the data transmitted by
the connected equipment.
Press the
key to reset the
error and then restart
communication.
<When the optional printer is in use>
This code appears when the call
number is “0” although the summary
setting for individual call numbers is
enabled.
Press the
key and set the
call number. Then restart
weighing.
<When the optional USB flash drive
feature is used>
This code appears when a USB flash
drive for writing is not connected.
Check that a USB flash drive
has been inserted.
Press the
key to reset the
error and then restart
communication.
<When the optional USB flash drive
feature is used>
This code appears when there is no
free space left in the USB flash drive
for writing.
Check the free space of the
USB flash drive.
Press the
key to reset the
error and then restart
communication.
<When the optional USB flash drive
feature is used>
This code appears when the USB
flash drive for writing has been set to
read only.
Check that the USB flash drive
is set to read/write.
Press the
key to reset the
error and then restart
communication.
<When the optional USB flash drive
feature is used>
This code appears when the writing
time into the USB flash drive is too
long.
Check the data to be written in
the USB flash drive.
Press the
key to reset the
error and then restart
communication.
Note:
Proper use of the instrument does not cause generation of errors.
If any of the above error codes is displayed, carry out the respective Corrective actions.
If the problem persists even after the above measures are taken, the instrument has internal
failure. Contact your dealer from which you purchased the instrument.
If “E-101” or higher number of error codes shows up, contact the dealer.
Содержание DP-5602