38
Operation Method
Program Creation Method
2
Edit step
Refer to the page 54 for the temperature rise/fall time.
They vary depending on the sample used or condition of exhaust opening (open or close). Check
them by conducting operation.
The conception of repeat is shown in the figure below. The first operation of repeat interval is not
counted as a repeat count.
Step No. S04
Start step: S02
Repeat times: 5
Step No. S01
No repeat
instruction
Step No. S02
No repeat
instruction
Step No. S03
No repeat
instruction
Step No. S05
No repeat
instruction
5 repeating
of interval
Program creation example;
The example here explains the program where the DNE wind speed adjustable device is equipped.
1
Program creation; Step 1 (S01)
TROUBLE
ENTER
CANCEL
DOOR
Select step
S01
Program OPR
Set TEMP
►
0
℃
S01
Set TEMP
►
0min
S01
①
Press the PROGRAM key. The program menu starts
and step number selection screen is displayed.
②
Press the ENTER key. The setting temperature
selection screen is displayed.
③
Press the ENTER key. The setting temperature edition
screen is displayed. Set "100
℃
" using the "
▼▲
" and
press the ENTER key to decide it.
④
Press the PROGRAM key. The setting period selection
screen is displayed.
⑤
Press the ENTER key. The setting period edition
screen is displayed. Set "25min" using the "
▼▲
" and
press the ENTER key to decide it.