Temperature output terminal
The temperature output signals for the Outlet (outlet temperature) and the Inlet (inlet temperature) are
4-20mA for the measure temperature of 40-320
°
C.
[Current output of 4-20mA:
Measured temperature of 40-320
°
C]
Conversion formula: Current output I (mA) = (0.05x measured temperature T(
°
C)
+
4)
Measured
temperature
T(
°
C) =20×current output I (mA)-80
When you connecting to the voltage input of the recorder, connect a fixed resistor (shunt resistor)
of 600
Ω
or lower to the voltage input of the recorder.
+
inlet temp
-
~20
40
)
+
-
(4 20mA: ~320 )
320
40
℃
(
mA
~
4
: ~
℃
outlet temp
20mA
4mA
0℃
320℃
Measured
temp.
Current output