YCO-1094-04-02
CPC-2000
Technical Manual
5
II Setting the Scale Functions
Calibrating the Scale
WARNING!
Calibration should only be performed by a qualifi ed scale tech ni cian. Beware of the haz-
ard of electric shock. Never work on powered circuits.
Enter the technician setup mode, select “END” and press the
N
G
key to enter the calibration pro-
gram. The scale will display raw counts. Press the
UNIT
key to select either pound or kilogram
calibration weights. With an empty platform on the load cell assembly, press the
TARE
key. The
scale should read “0”. If “0” is not displayed, then press the
TARE
key again until “0” is displayed.
Place either 2 kg or 5 lb (depending on the units selected above) on the scale platform. Wait for the
counts to sta bi lize and press the
TARE
key again. The scale should briefl y indicate either 2 kg or
5 lb and then power off. Turn the scale back on and verify that it is reading accurately. If not, then
repeat the calibration procedure.