![Yamato CA300 Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/yamato/ca300/ca300_instruction-manual_905292017.webp)
13
Handling Precautions
WARNING!
If a problem occurs
If smoke or strange odor should come out of this unit for some reason, turn off the power key
right away, and then turn off the breaker and the main power. Immediately contact a service
technician for inspection. If this procedure is not followed, fire or electrical shock may result.
Never perform repair work yourself, since it is dangerous and not recommended.
Measure for flammability and handling of flammable solvent
This unit is not designed as the explosion-proof construction. Pay special attention to the
handling of the sample to be handled with this unit on the consumption with the explosive
material, flammable material, and similar ones. The flammable material may be vaporized by
leaving it at the temperature higher than room temperature, and could cause the fire or
explosion. When handling such material, provide ventilation with enough before the operation.
(Refer to page22 “List of Dangerous Substances”.)
CAUTION!
Maximum trap capacity
The max. trap capacity for the water type liquid is around 0.9kg (CA300) or 0.8kg (CA800).
Inserting too much trap to this unit could cause the clogging of "IN", "OUT" nozzle.
Do not step on this unit
Do not step on this unit. It will cause injury if this unit fall down or break.
Do not put anything on this unit
Do not put anything on this unit. It will cause injury if fall.
During a thunder storm
During a thunderstorm, turn off
the power key immediately, then turn off the circuit breaker and
the main power. If this procedure is not followed, fire or electrical shock may be caused.
Countermeasure for stop operation during night or long-term stop
In case of stopping operation during night or long-term, toggle the power switch to "OFF".