15
4. OPERATION PROCEDURES
Operation Modes & Functions
BM/BO series operation modes are outlined in the table below:
No.
Name
Description
Page
1
Constant Temp Mode
Set desired temperature with
△ ▽
. Press and hold
START/STOP to start operation. Press START/STOP again to
stop operation.
18
2
Quick Auto Stop
Mode
Press TIMER key while constant temp mode is running. Select
and set stop timer using
△ ▽
. Press and hold
START/STOP to start quick auto stop mode.
3
Auto Stop Mode
Press TIMER key while unit is in standby. Select
and
set operation time using
△▽
. Press and hold START/STOP to
start auto stop mode.
21
4
Auto Start Mode
Press TIMER key twice while unit is in standby. Select
and set operation time using
△ ▽
. Press and hold
START/STOP to start auto start mode.
BM/BO series safety features are outlined in the table below:
No.
Name
Description
Page
1
Overheat Prevention
Devices
①
Overheat prevention function:
This function is set to automatically activate (manual
reset) when actual temperature exceeds setting by
40°C.
[Er06] appears in the display screen. Turn off the main
power switch and call for service.
Unit recovers when problem source is eliminated and
unit is restarted.
25
②
Thermal fuse:
Unit contains a thermal fuse which will cut power to the
heater in the event that the overheat prevention function
mentioned above fails to operate.
2
Main Power Switch
Circuit breaking power switch located on the rear panel of unit.
-
Содержание BM500