24
4. OPERATION PROCEDURES
Building Programs
⑨
Press the ENTER key.
, indicating timer setting
for step 1, will be shown in the top display. Current timer
setting will also be shown flashing in the center display.
Before setting the timer, be sure to confirm temperature unit
rise/fall capability.
For example, about 30 minutes is needed for a ±10
C
increase from 50
C to 60
C (see tables on pg. 21 above),
on the BA500 model, which is about 1
C every 6 minutes.
Accordingly, it would take approximately 60 minutes to
reach 70
C at the same rate. This does not include
temperature stabilization time.
Maximum timer setting for each step is 999 hours and 50
minutes.
⑩
When timer is set, press the ENTER key. "Sv_2",
indicating temperature for step 2 can now be set, will be
shown in top display. Enter temperature and time using
the same procedure, decribed thus far, for all steps (use the
program planning worksheet on pages 27 & 28).
⑪
When the repeat function becomes necessary, press the
SUBMENU key after setting timer (step 7 in the above
example) in the step where repeat operation is to be used.
This brings up repeat function setting mode.
Follow Pattern Repeat Function (P.26) for program repeat
function entry procedure.
⑫
Screen returns to the initial settings screen, after setting
temperature and timer in the final step.
Verification run:
When possible, confirm temperatures and times in a newly
entered program by running program with unit unloaded once,
before using program on actual test samples.
5. Run program
Press the START/STOP key for about one second. Program
will begin running.
PROGRAM lamp illuminates and the top display will show
, signifying that step 1 is currently under way.
Use the
▼ to monitor temperature and time (top display)
remaining in a currently running step.