Yamatake Corporation
Operation
Model HGC303 - Heat Value Gas Chromatograph
3-27
Trend graph of carrier gas pressure and oven temperature control
Figure 3-11 Trend graph of carrier pressure and oven temp. control
This data is displayed according to the set data saving interval (Refer to “3-3-4 : Set up
HGM” on page 3-15).
Default interval: 5 sec.
* In case of Windows 98 or Me, data are also saved in C:\My documents\MyHGC.
Table 3-14 Description of trend graph of carrier gas pressure and oven temperature control
No.
Display
Description
1
URV (Over Temp.)
Upper Range Value for oven temperature.
Click on the URV button to change the URV value
Default value: 62 deg.C.
2
Load (Offline)
Recall saved data. File name extension:.sv1
3
Save (Online)
The latest data is saved.
Default directory is “C:\Program files\hgm\data”.*
4
Verify that your printer is connected and working properly.
5
30min
(Time range select)
This indicates the time range of the horizontal axis. Select a time
range from pull-down menu:30 min., 60 min., 3 hours, 6 hours, 12
hours, 1 day, 2 days, 3 days, 6 days, 12 days
Default: 30min.
6
LRV (Oven Temp.)
Lower Range Value for oven temperature
Default value: 54 deg. C.
Click on the LRV button to change the LRV value
7
URV (Carr Press.)
Upper Range Value for carrier pressure
Default value: 350kPa
Click on the URV button to change the URV value
8
LRV (Carr Press.)
Lower Range Value for carrier pressure.
Default value: 150 kPa
Click on the LRV button to change the LRV value
9
Black circle
Black indicates oven temperature.
10
Red circle
Red indicates carrier pressure.
11
Time
Online: Date and time of the latest data
Offline: Date and time of when the data was saved.
1
2
3
4
5
6
8
9
10
11
7
HGM000019000S
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...