Appendix
Yamatake Corporation
A-2
Model HGC303 - Heat Value Gas Chromatograph
* GHV: Gross Heating Value (British Thermal Unit / Cubic Foot)
Decide which normalization method should be selected, and configure it to the model
HGC303 at the configuration mode. In addition, don't forget to setup it at the [Setup
HGM] screen for HGM.
Next describes how to calculate heating value and other properties.
Table A-1 An example of the calculated normalized concentration and heating
value
PV
Component name
Un- normalized
conc. (yi mol%)
Normalized conc. (mol%)
Standard
Methane
1
C6+
0.05
0.05
0.05
2
C3H8
1.5
1.47
1.5
3
i-C4H10
0.3
0.29
0.3
4
n-C4H10
0.3
0.29
0.3
5
neo-C5H12
0.1
0.1
0.1
6
i-C5H12
0.05
0.05
0.05
7
n-C5H12
0.05
0.05
0.05
8
N2
3
2.95
3
9
CH4
93
91.31
91.15
10
CO2
0.5
0.49
0.5
11
C2H6
3
2.95
3
Sum
101.85
100
100
GHV (BTU/CF)*
1045.86
1046.44
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...