Matrix adjustment, check and calibration
35
ba76207e02
12/2020
be valid, the potential level, slope and drift voltage must be within
certain ranges.
Valid ranges for slope and drift potential:
The valid potential levels (MIN, MAX) can be taken from the online help.
Taking over the
determined values
For each successfully calibrated electrode you can individually decide
whether the values should be stored for measurement.
Sensor history
The result of the calibration is available in the
Sensor history
(see
).
4.4
Sensor history
The data of the calibration procedures and matrix adjustments are
available in the Sensor history and in the calibration history.
The sensor history supplements and expands the calibration history:
The results of any checks are displayed.
All data of all electrodes of the sensor can be displayed in one list.
More datasets can be displayed for each electrode.
Sensor history
1
In the measured value display, select the sensor with
<
>
and
use
<C>
to open the
PROCEDURE/SETTINGS
window for the
sensor.
2
Select the menu
Sensor history
.
3
Select the electrodes that should be displayed in the
Sensor
history
.
4
Follow the user guide.
The
Sensor history
is displayed for the selected electrodes.
Value of the slope:
50 ... 70 mV
Drift voltage:
-45 ... +45 mV
Actions for error elimination are given in chapter 8 W
....
Using
<S>
, you can save the displayed Sensor history to a
USB memory device connected (directory, VARiON).