TruLab pH 1110
ORP
ba76155e01
03/2014
27
6
ORP
6.1
Measuring
6.1.1
Measuring the ORP
1.
Connect the ORP electrode to the meter.
2.
If necessary, call up the measured parameter U with
<M>
.
3.
Immerse the ORP electrode in the test sample.
The measured value is checked for stability (stability control).
The
[AR]
status display flashes.
4.
Wait for a stable measured value.
The
[AR]
display indicator no longer flashes.
Stability control (AutoRead )
During the measuring procedure, the stability control function is automatically
activated.
The stability control function (
AutoRead
) continually checks the stability of the
measured values in the monitored time interval. The stability has a consider-
able impact on the reproducibility of measured values. The
[AR]
display indi-
cator flashes until a stable value is measured.
Stability criteria (AutoRead )
3
6
1
8
2
4
U
°C
TP
mV
AR
3
Measured parameter
Time interval
Stability in the time interval
ORP
15 seconds
∆
: Better than 0.3 mV
Temperature
15 seconds
∆
: Better than 0.5 °C