FM/PC, FM-Case/PC, Filter/PC
Installing the filter module
7
ba76171e02
04/2019
NOTE
Make sure that the frame is clean for the mounting. If solid parti-
cles such as sand grains are between the frame and filter plate,
the membrane could be damaged when the frame parts are
screwed together.
Disassembly
To disassemble the frame and filter plate, proceed in reverse order of
installation.
4
Installing the filter module
Bild 4-1
Filtration unit (filter module mounted on the guide slide)
1
Fix the filter module to the height adjustable guide slide (scope of
delivery: suction tube) with the aid of 2 Allen screws (Allen
wrench size 4). The tube connections must point downward.
2
Connect the blue sample tube to the quick coupling of the adapter
of the filter plate. The adapter (see pos. 2 in Bild 3-1) is freely
rotatable.
3
Submerse the filtration unit in the basin or channel.
The frame only serves as a bracket, not as a seal. The fil-
ter plate is tight itself.
The blue sample tube must be clean, especially where it
should be connected.
Dirt particles in the sample flow can, for example, block the
valves.