![Xylem YSI Filter/PC Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/xylem/ysi-filter-pc/ysi-filter-pc_mounting-instructions_892052008.webp)
Maintenance / cleaning
FM/PC, FM-Case/PC, Filter/PC
8
ba76171e02
04/2019
4
Further proceeding as described in the operating manual of your
analyzer (chapter C
OMMISSIONING
).
Bild 4-2
Filtration unit (filter module mounted on the guide slide)
5
Maintenance / cleaning
5.1
Cleaning the filter plate
When to clean the filter plate can be determined empirically based on
the negative pressure display on the filtration pump.
With a delivery height of approx. 2 m and a new filter plate, the neg-
ative pressure is approx. -0.3 bar.
Every meter of delivery height increases the negative pressure by