Commissioning, measuring

FDO 4410(W)

8

ba76197e01

10/2017

2.4

Measuring

Minimum depth of

immersion

Observe the required minimum immersion depth (see chapter 5 T

ECHNICAL

 

DATA

).

Incident flow

The FDO 4410(W) D.O. sensor enables precise measurements without any 
incident flow.

However, an incident flow of the sensor membrane improves the responding 
behavior of the sensor (see chapter 5 T

ECHNICAL

 

DATA

). The incident flow can 

be provided in different ways, e. g.:

The flow of the water to be measured is sufficient (aeration tank, water 
pipe, stream)

Slowly pull the sensor through the water by hand (lake, container), or

Use a flow aid such as a magnetic stirrer with stirring device (see 
chapter 6 W

EAR

 

PARTS

 

AND

 

ACCESSORIES

)

2.5

Function check and user calibration

Factory calibration

The FDO 4410(W) is factory calibrated. In the recommended application (see 
page 5), the measuring characteristics of the sensor cap remain stable for the 
specified service life. Thus, a user calibration is not usually required.

When does a

check or user

calibration make

sense?

A function check or user calibration can be useful in the following special 
cases:

If the measured values appear to be implausible and it is assumed that the 
service life of the sensor cap is over

Routinely within the framework of the company quality assurance.

Check- or

calibration

medium

The check and user calibration take place in water vapor-saturated air. The 
suitable conditions are easily provided with the aid of the check and calibra-
tion vessel, 

FDO Check

. For this purpose, moisten the sponge inside the ves-

sel. Then insert the sensor in the vessel as far as it will go. The sensor 
membrane must be clean and dry for this.

Moisten the sponge:

Remove the cap (1).

Take out the sponge (2), wet it, then slightly squeeze it out.

Insert the sponge again and close the calibration and storage vessel with 

1

2

Содержание YSI FDO 4410

Страница 1: ...OPERATIONS MANUAL ba76197e01 10 2017 FDO 4410 FDO 4410W OPTICAL D O SENSOR FDO 4410 FDO 4410W OPERATING MANUAL...

Страница 2: ...2 ba76197e01 10 2017 FDO 4410 Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email environmental ysi com Internet www ysi com Copyright 2017 Xylem Inc...

Страница 3: ...pening and closing the IDS plug connection FDO 4410W 7 2 4 Measuring 8 2 5 Function check and user calibration 8 2 6 Storage 9 3 Maintenance cleaning replacement 10 3 1 General maintenance instruction...

Страница 4: ...tronics 3 Connection cable FDO 4410 4 IDS plug FDO 4410W 5 Protective hood 6 Sensor membrane 7 SC FDO 4410 sensor cap with memory chip 8 Gold plated contacts for the memory chip of the sensor cap 9 Lo...

Страница 5: ...ensor is connected and is used for measurement and for measured value documentation The calibration data are stored in the sensor cap Thus the calibration is automatically retained if the sensor or me...

Страница 6: ...ure scratches 2 2 Commissioning Scope of delivery D O sensor FDO 4410 W with sensor cap Check and storage vessel FDO Check Operating manual Commissioning Connect the electrode to the meter FDO 4410 vi...

Страница 7: ...ng device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plug connecti...

Страница 8: ...d application see page 5 the measuring characteristics of the sensor cap remain stable for the specified service life Thus a user calibration is not usually required When does a check or user calibrat...

Страница 9: ...for the temperatures of the sensor and cali bration vessel to adjust 2 6 Storage Always store the sensor in the check and storage vessel FDO Check at a temperature in the range 0 50 C 32 122 F The ste...

Страница 10: ...clean and dry prior to removing the sensor cap Please do not touch the outer sensor membrane with your fingers Touch the sensor cap at the sides only arrows in figure on the left Avoid any great mech...

Страница 11: ...Do not push any tools or other sharp objects between the sealing sur faces This would damage the sealing surfaces and sealing ring 6 Check the front surface of the sensor for absolute cleanness and c...

Страница 12: ...h In the case of persisting dirt you can add some household washing up liq uid to the tapwater Attention Never use any alcohol for cleaning 3 3 2 Interior cleaning of sensor cap and sensor head If moi...

Страница 13: ...check the zero point of the sensor Measurement in a nitrogen atmosphere recommended method Measurement in a sodium sulfite solution according to DIN EN 25814 ISO 5814 Test criterion The sensor is OK...

Страница 14: ...n 3 2 Dirt inside the sensor cap and sensor head Protective hood not properly tightened Sensor cap untight or defective Dismantle the sensor cap Clean the insides of the sensor cap and sensor head see...

Страница 15: ...25 C 77 F Dimensions in mm Weights FDO 4410 180 g with 3 m cable FDO 4410W 107 g FDO Check 58 g Materials Shaft POM Connection head POM Sensor cap PVC silicone and PMMA Sensor head POM PV and PMMA Se...

Страница 16: ...istance Sensor with connection cable IP 68 1 x 106 Pa or 10 bar Cable plug FDO 4410 IP 67 when plugged in The FDO 4410 W meets the requirements according to article 3 3 of the 97 23 EC directive Press...

Страница 17: ...bar O2 partial pressure Response time at 20 C 68 F in stirred solution t90 90 of the final value display after 30 s t95 95 of the final value display after 45 s t99 99 of the final value display after...

Страница 18: ...sensor cap SC FDO 4410 201 310Y Connection cable FDO 4410 W meter Description Model Order no IDS connection cable 1 5 m IDS CABLE 1 5 903 850Y IDS connection cable 3 m IDS CABLE 3 903 851Y Radio conn...

Страница 19: ...uct for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Product Return for...

Страница 20: ...Contact Information FDO 4410 20 ba76197d01 10 2017...

Страница 21: ......

Страница 22: ...and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with cus...

Отзывы: