Commissioning
TriOxmatic
®
690
8
ba25314e05
09/2020
3
Commissioning
3.1
Scope of delivery
TriOxmatic
®
690
The sensor is filled with electrolyte and provided with a protective
hood
and protective caps.
ZBK 600 accessory kit
Operating manual.
3.2
Installation
Connection to the
measuring transmitter
The TriOxmatic
®
690 connection cable has a screw plug to be
connected to the Oxi input socket of a measuring transmitter. It is
connected to the terminal strip of measuring transmitters without an Oxi
input socket via the ADA/AMPH adapter. For detailed information
please refer to the operating manual of the measuring transmitter.
3.3
Commissioning / Getting the instrument ready for
measuring
Precision
measurements
Recommendation: In order to carry out precision measurements, leave
the sensor to polarize for a longer period of time, e.g. overnight, and
recalibrate it the following day.
Do not suspend the sensor on the sensor connection cable.
Use a sensor holder or armature. Information on this and
other TriOxmatic
®
690 accessories is given in the WTW
catalog and on the Internet.
1
Pull the protective cap off the protective hood of the sensor.
2
Leave the sensor to lie in the air for at least 60 minutes while it
is switched on (polarization).
For the sensor to polarize, the following conditions must be
fulfilled:
The sensor is connected with the measuring transmitter.
The measuring transmitter is in operation
3
Calibrate the sensor (see operating manual of the measuring
transmitter).