Cleaning

Remove water-soluble contamination by rinsing with deionized water. 
Remove other contamination as follows:

If you want to clean and rinse the ground junction, you can open it as follows: 

completely unscrew the ground stopper (see figure on page 5).

After cleaning

Rinse the electrode with deionized water and condition it in reference 
electrolyte solution for at least 1 hour. Then recalibrate the electrode.

6

Technical data

Contamination

Cleaning procedure

Fat and oil

Rinse with water containing household 
washing-up liquid

Lime and hydroxide depos-
its

Rinse with citric acid (10 % by weight)

Proteins

Immerse in pepsin cleaning solution PEP/pH 
for approx. 1 hour. 
Note: Make sure the level of the reference 
electrolyte is above that of the cleaning solu-
tion.

Hydrofluoric acid, hot phosphoric acid and strong alkaline 
solutions destroy the glass membrane.

Measurement

pH measuring range

0.000 ... 14.000

Allowed temperature range

0 ... 60 °C

Содержание WTW SenTix HW-T 900

Страница 1: ...OPERATING MANUAL ba77090e03 06 2021 SenTix HW T 900 P pH ELECTRODE WITH LIQUID REFERENCE SYSTEM AND GROUND JUNCTION OPERATING MANUAL...

Страница 2: ...SenTix HW T 900 P 2 ba77090e03 06 2021 Copyright 2021 Xylem Analytics Germany GmbH Printed in Germany...

Страница 3: ...d the last 10 calibrations The data is recalled by the meter when the sensor is connected and is used for measurement and for measured value documentation Storing the calibration data in the sensor en...

Страница 4: ...Open the locking device step 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the...

Страница 5: ...ctrode according to the operating manual of the meter When doing so observe the general rules on the following page Ground junction of SenTix HW T 900 P With the ground junction of the SenTix HW T 900...

Страница 6: ...nt of any solution sample or buffer solution from one measurement to the next by taking the following measures Shortly rinse the calibration and sample beakers with the solution the Special wrench Gro...

Страница 7: ...rage During short measuring breaks With the refilling opening open immerse the electrode in reference electrolyte KCl 3 mol L Ag free Observe the minimum depth of immersion Prior to the next measureme...

Страница 8: ...H and temperature The warranty does not cover failure caused by measuring conditions and mechanical damage 5 Maintenance and cleaning Refilling the reference electrolyte During operation a small amoun...

Страница 9: ...electrolyte solution for at least 1 hour Then recalibrate the electrode 6 Technical data Contamination Cleaning procedure Fat and oil Rinse with water containing household washing up liquid Lime and h...

Страница 10: ...mm Smallest allowed bend radius Fixed installation 20 mm Flexible use 60 mm Plug type Socket 4 pins Shaft dimensions material Shaft length 170 mm Shaft diameter 12 mm Shaft material Glass IDS plug Syn...

Страница 11: ...n Model Order no IDS connection cable 1 5 m AS IDS 1 5 903 850 IDS connection cable 3 m AS IDS 3 903 851 IDS connection cable 6 m AS IDS 6 903 852 IDS connection cable 10 m AS IDS 10 903 853 IDS conne...

Страница 12: ...Disposal SenTix HW T 900 P 12 ba77090e03 06 2021...

Страница 13: ......

Страница 14: ...rove the way water is used conserved and re used in the future is central to our work Our products and services move treat analyze monitor and return water to the environment in public utility industr...

Отзывы: