Measuring / Operation

Note

Calibrate the measuring system after the initial commissioning and at 
regular intervals (depending on the application).

4.1

Calibration

Why calibrate?

During the operation of a pH electrode, the slope and asymmetry of the 
electrode changes with time. The calibration procedure determines the 
current slope and asymmetry of the electrode.

When to calibrate?

Calibrate before measuring and at regular intervals (depending on the 
application). 

Calibration procedures

The available calibration procedures depend on the measuring 
transmitter used. The individual steps for the calibration are 
comprehensively described in the operating manual of the measuring 
transmitter.

4.2

Measuring

Warning

Contact with the sample can lead to danger to the user! 
Depending on the type of sample, suitable protective measures 
must be taken (protective clothing, protective goggles, etc.).

Please pay attention to:

the minimum immersion depth of the armature (> 40 mm)

the measuring range of the electrode used (see operating manual of 
the electrode).

Содержание WTW SensoLyt 700

Страница 1: ...OPERATING MANUAL ba15307e05 09 2020 SensoLyt 700 PH ORP ARMATURE WITH IMPEDANCE CONVERTER AND TEMPERATURE SENSOR SensoLyt 700 SensoLyt 700 SW ...

Страница 2: ...2 ba15307e05 09 2020 SensoLyt 700 Copyright 2020 Xylem Analytics Germany GmbH Printed in Germany ...

Страница 3: ...Installation 8 3 3 Commissioning Getting the instrument ready for measuring 9 4 Measuring Operation 12 4 1 Calibration 12 4 2 Measuring 12 5 Maintenance and changing the electrode 13 5 1 General maintenance instructions 13 5 2 Replacing the combination electrode 13 5 3 Cleaning 17 5 4 Disposal 17 6 Replacement parts and accessories 18 6 1 Combination electrodes 18 6 2 General accessories 18 7 What...

Страница 4: ... even over greater distances Glass breakage monitoring The sensor is equipped with a SensCheck function for monitoring glass breakage 1 2 Recommended fields of application In conjunction with the SensoLyt SEA SensoLyt DWA and SensoLyt ECA pH combination electrodes as well as the SensoLyt PtA ORP combination electrode the SensoLyt 700 pH ORP armature is suitable for stationary pH or ORP measurement...

Страница 5: ...rmature General safety instructions Safety instructions in this operating manual can be recognized by the warning symbol triangle in the left column The signal word e g Caution indicates the level of the danger Warning indicates instructions that must be followed precisely in order to prevent serious dangers to persons Caution indicates instructions that must be followed precisely in order to avoi...

Страница 6: ...ersonnel are familiar with the necessary precautions to take when dealing with chemicals as a result of their professional training and experience 2 4 General safety instructions Function and operational safety The armature left the factory in a safe and secure technical condition The failure free function and operational safety of the armature is only guaranteed if the generally applicable safety...

Страница 7: ...es as described in this manual If you are in any doubt contact the supplier of your armature Obligations of the operator The operator of the armature must ensure that the following rules and regulations are followed when dealing with hazardous substances EEC directives for protective labor legislation National protective labor legislation Safety regulations Safety data sheets of the chemical manuf...

Страница 8: ...er with the SensoLyt 700 pH ORP armature is given in section 6 1 COMBINATION ELECTRODES Connection to the measuring transmitter The connection cable of the SensoLyt 700 is ready to be connected to the terminal strip of a measuring transmitter with high impedance pH ORP input For all further information please refer to the operating manual of the measuring transmitter The assignment of the cable wi...

Страница 9: ...p of the sensor to keep the electrode active during storage or during longer pauses in measuring The cap must be removed for measuring Mounting the combination electrode 1 Unscrew the protective hood from the armature 2 Pull off the blind plug from the plug head socket of the armature 3 Screw off the protective cap of the plug head connector of the electrode ...

Страница 10: ...nected electrode into the armature right up to the stop so that the connection is watertight 4 Screw the electrode into the plug head socket of the armature 5 Push the unit into the armature up to the stop 6 Pull the KCI filled plastic cap off the electrode for measuring ...

Страница 11: ...missioning 11 ba15307e05 09 2020 7 Screw the protective hood onto the armature 8 Make the settings for the electrode on the measuring transmitter Subsequently calibrate the measuring system see section 4 1 CALIBRATION ...

Страница 12: ...efore measuring and at regular intervals depending on the application Calibration procedures The available calibration procedures depend on the measuring transmitter used The individual steps for the calibration are comprehensively described in the operating manual of the measuring transmitter 4 2 Measuring Warning Contact with the sample can lead to danger to the user Depending on the type of sam...

Страница 13: ...t operating manual of the electrode 5 2 Replacing the combination electrode Warning Contact with the sample can lead to danger to the user Depending on the type of sample suitable protective measures must be taken protective clothing protective goggles etc Caution If the glass of the pH electrode breaks there is a danger of cuts from the splinters of glass If it is necessary to replace an electrod...

Страница 14: ... ba15307e05 09 2020 2 Use the protective hood as a tool to lever out the electrode 3 Carefully pull out the electrode until the plug head screwed fitting can be seen 4 Unscrew the combination electrode from the plug head socket for disposal see section 5 4 ...

Страница 15: ... the connected electrode into the armature right up to the stop so that the connection is watertight 5 Screw in a new electrode 6 Push the unit into the armature up to the stop 7 Pull the KCI filled plastic cap off the electrode for measuring 8 Screw the protective hood onto the armature ...

Страница 16: ...Maintenance and changing the electrode SensoLyt 700 16 ba15307e05 09 2020 9 Calibrate the measuring system see section 4 1 CALIBRATION ...

Страница 17: ...e function check Exterior cleaning Note Please also observe the instructions on cleaning the electrode in the respective operating manual 5 4 Disposal Armature We recommend disposing of the armature as electronic refuse Combination electrodes If no official regulations apply to the contrary used and defective electrodes can be treated as household waste Contamination Cleaning agents For sludge and...

Страница 18: ...ble for the pressure range and temperature range planned pH combination electrodes ORP combination electrode 6 2 General accessories Technical buffer solutions for pH calibration Protective hood Note Information on further accessories is given in the WTW catalog and on the Internet Model Order no SensoLyt SEA SensoLyt DWA SensoLyt ECA SensoLyt SEA HP 109 115 109 119 109 117 109 118 SensoLyt PtA 10...

Страница 19: ...rode connection Watering cap still on the electrode Pull off watering cap and calibrate No or wrong calibration performed Calibrate Electrode contaminated Clean electrode Liquid has penetrated the armature Armature defective return to WTW Instrument setting incorrect Correct instrument setting Cause Remedy Slope of the electrode too low Replace electrode Asymmetry of the electrode too high Replace...

Страница 20: ...e resistance Max allowed overpressure sensor including connection cable The SensoLyt 700 meets the requirements according to article 3 3 of the directive 97 23 EC pressure equipment directive Sensor accuracy 0 3 K Response time t99 99 of the final value display after 15 s Measuring medium 0 C 60 C 32 140 F Storage transport 5 C 65 C 23 149 F with installed combination electrode SensoLyt SEA DWA Pt...

Страница 21: ...oring of the pH electrode by the measuring transmitter with installed combination electrode SensoLyt SEA DWA PtA min 40 mm max 100 m or to cable end with installed combination electrode SensoLyt ECA min 40 mm max 60 m or to cable end with installed combination electrode SensoLyt SEA HP min 40 mm max 100 m or to cable end temperature dependent see safety instruction below in the entire temperature ...

Страница 22: ...mbination electrode Electrodes that can be integrated Connection technique Connection cable permanently mounted on the sensor Connection to the measuring transmitter via 7 pole screw plug SensoLyt 700 approx 320 g SensoLyt 700 SW approx 880 g pH combination electrodes SensoLyt SEA SEA HP DWA ECA ORP electrodes SensoLyt PtA ...

Страница 23: ...le POM Temperature sensor SensoLyt 700 SensoLyt 700 SW V4A stainless steel 1 4571 V4A stainless steel 1 4571 KTL coated Closing head POM Protection ring POM Cable gland SensoLyt 700 SensoLyt 700 SW V4A stainless steel 1 4571 Titan Grade 2 Cable sheath PUR Length SensoLyt 700 SensoLyt 700 SW 7 m special lengths on request 15 m Diameter 8 6 mm Smallest allowed bend radius Permanent bend 130 mm Short...

Страница 24: ...700 24 ba15307e05 09 2020 8 4 Electrical data Pin assignment Pin Assignment Voltage Max current 1 2 3 4 5 6 7 Ua output signal 0 V Ub Ub NTC NTC 10 V 10 V 3 5 V 3 5 V 1 mA 1 mA 0 15 mA 0 15 mA 4 3 5 6 7 1 2 Plug from the front ...

Страница 25: ......

Страница 26: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise backed by a legacy of innovation Xylem Analytics Germany GmbH Dr Karl Slevogt Str 1 82362 Weilheim Germany Xylem Analytics Germany Sales GmbH Co KG WTW Dr Karl Slevogt Str 1 82362 Weilheim Ge...

Отзывы: