Settings
pH/Cond 3320
52
ba77054d01
02/2014
Application tips
Set the temperature compensation suitable for your test sample:
8.1.8
Setting the TDS factor
The factor to calculate the total dissolved solids is set to 1.00 in the delivery
condition.
You can adjust this factor to meet your requirements in the range 0.40 ... 1.00.
The factor is set in the menu for the parameter TDS (
Measurement
/
TDS fac-
tor
).
The reference temperature and temperature compensation are set
in the
Measurement
menu for the parameter, conductivity (see sec-
Test sample
Temperature compensation
Display
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water
nLF
according to EN 27 888
nLF
Other aqueous solutions
Lin
Set linear
temperature coefficient
0.000 ... 10.000 %/K
Lin
Salinity (seawater)
Automatic
nLF
according to
IOT (International Oceano-
graphic Tables)
Sal
,
nLF